Monitoring of recruitment and derecruitment by electrical impedance tomography in a model of acute lung injury
- 1 December 2000
- journal article
- Published by Wolters Kluwer Health in Critical Care Medicine
- Vol. 28 (12) , 3891-3895
- https://doi.org/10.1097/00003246-200012000-00025
Abstract
OBJECTIVE: To evaluate a noninvasive system for obtaining information about alveolar recruitment and derecruitment in a model of acute lung injury. DESIGN: Prospective experimental study. SETTING: Animal research laboratory. SUBJECTS: Nine anesthetized pigs. INTERVENTIONS: Electrical impedance tomography measurements were performed. Electrical impedance tomography is an imaging technique that can register the ventilation-induced impedance changes in different parts of the lung. In nine anesthetized pigs, repeated lung lavages were performed until a PaO2 of <80 mm Hg was reached. Thereafter, the lungs were recruited according to two different recruitment protocols: the open lung approach and the open lung concept. Five time points for measurements were chosen: healthy (reference), lavage (atelectasis), recruitment, derecruitment, and maintain recruited (final). MEASUREMENTS AND MAIN RESULTS: After lavage, there was a significant increase in the impedance ratio, defined as the ventilation-induced impedance changes of the anterior part of the lung divided by that of the posterior part (from 1.75 +/- 0.63 to 4.51 +/- 2.22; p <.05). The impedance ratio decreased significantly after performing the recruitment protocol (from 4.51 +/- 2.22 to 1.18 +/- 0.51). During both recruitment procedures, a steep increase in baseline impedance change was seen. Furthermore, during derecruitment, a decrease in the slope in baseline impedance change was seen in the posterior part of the lung, whereas the anterior part showed no change. CONCLUSION: Electrical impedance tomography is a technique that can show impedance changes resembling recruitment and derecruitment of alveoli in the anterior and posterior parts of the lung. Therefore, electrical impedance tomography may help in determining the optimal mechanical ventilation in a patient with acute lung injurKeywords
This publication has 12 references indexed in Scilit:
- Electrical impedance tomography in monitoring experimental lung injuryIntensive Care Medicine, 1998
- Effect of a Protective-Ventilation Strategy on Mortality in the Acute Respiratory Distress SyndromeNew England Journal of Medicine, 1998
- Beneficial effects of the "open lung approach" with low distending pressures in acute respiratory distress syndrome. A prospective randomized study on mechanical ventilation.American Journal of Respiratory and Critical Care Medicine, 1995
- Changes in the thoracic impedance distribution under different ventilatory conditionsPhysiological Measurement, 1995
- Effects of positive end-expiratory pressure on regional distribution of tidal volume and recruitment in adult respiratory distress syndrome.American Journal of Respiratory and Critical Care Medicine, 1995
- A real-time electrical impedance tomography system for clinical use-design and preliminary resultsIEEE Transactions on Biomedical Engineering, 1995
- Open up the lung and keep the lung openIntensive Care Medicine, 1992
- Applications of applied potential tomography (APT) in respiratory medicineClinical Physics and Physiological Measurement, 1987
- Pressure-Volume Curve of Total Respiratory System in Acute Respiratory Failure: Computed Tomographic Scan StudyAmerican Review of Respiratory Disease, 1987
- In Vivo Lung Lavage as an Experimental Model of the Respiratory Distress SyndromeActa Anaesthesiologica Scandinavica, 1980