Evaluation of Automated Sample Preparation and Quantitative PCR LCx Assay for Determination of Human Immunodeficiency Virus Type 1 RNA
Open Access
- 1 April 2004
- journal article
- research article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 42 (4) , 1439-1443
- https://doi.org/10.1128/jcm.42.4.1439-1443.2004
Abstract
Efforts have been made to achieve full automation of molecular assays for quantitative detection of human immunodeficiency virus type 1 (HIV-1). In the present study, the Abbott LCx HIV RNA Quantitative assay was evaluated in conjunction with automated HIV-1 RNA extraction on the MagNA Pure LC instrument and compared to the conventional LCx HIV RNA Quantitative assay, which uses a manual nucleic acid extraction protocol. Accuracy, linearity, and interassay and intra-assay variations were determined. The performance of the assay in a routine clinical laboratory was tested with a total of 105 clinical specimens. When the accuracy of the LCx HIV RNA Quantitative assay with the automated sample preparation protocol was tested, all results were found to be within ±0.5 log unit of the expected results. Determination of linearity resulted in a quasilinear curve over 3.5 log units. For determination of interassay variation, coefficients of variation were found to be between 21 and 66% for the LCx HIV RNA Quantitative assay with the automated sample preparation protocol and between 10 and 69% for the LCx HIV RNA Quantitative assay with manual sample preparation. For determination of intra-assay variation, coefficients of variation were found to be between 7 and 25% for the LCx HIV RNA Quantitative assay with the automated sample preparation protocol and between 7 and 19% for the LCx HIV RNA Quantitative assay with manual sample preparation. When clinical samples were tested by the LCx HIV RNA Quantitative assay with the automated sample preparation protocol and the results were compared with those of the LCx HIV RNA Quantitative assay with manual sample preparation, 95% of all positive results were found to be within ±0.5 log unit. In conclusion, the assay with automated sample preparation proved to be suitable for use in the routine diagnostic laboratory and required significantly less hands-on time.Keywords
This publication has 28 references indexed in Scilit:
- Entirely Automated Quantification of Human Immunodeficiency Virus Type 1 (HIV-1) RNA in Plasma by Using the Ultrasensitive COBAS AMPLICOR HIV-1 Monitor Test and RNA Purification on the MagNA Pure LC InstrumentJournal of Clinical Microbiology, 2003
- Diagnosing Herpesvirus Infections by Real-Time Amplification and Rapid CultureJournal of Clinical Microbiology, 2003
- Qualitative detection of Legionella species in bronchoalveolar lavages and induced sputa by automated DNA extraction and real-time polymerase chain reactionMedical Microbiology and Immunology, 2002
- Detection of Methicillin-ResistantStaphylococcus aureusand Simultaneous Confirmation by Automated Nucleic Acid Extraction and Real-Time PCRJournal of Clinical Microbiology, 2002
- Evaluation of an Automated Sample Preparation Protocol for Quantitative Detection of Hepatitis C Virus RNAJournal of Clinical Microbiology, 2002
- Rapid detection of enterovirus infection by automated RNA extraction and real-time fluorescence PCRJournal of Clinical Virology, 2002
- Determination of Human Immunodeficiency Virus Type 1 Subtypes by a Rapid Method Useful for the Routine Diagnostic LaboratoryClinical and Diagnostic Laboratory Immunology, 2001
- Detection of Herpes Simplex Virus DNA in Genital and Dermal Specimens by LightCycler PCR after Extraction using the IsoQuick, MagNA Pure, and BioRobot 9604 MethodsJournal of Clinical Microbiology, 2001
- Single Rapid Real-Time Monitored Isothermal RNA Amplification Assay for Quantification of Human Immunodeficiency Virus Type 1 Isolates from Groups M, N, and OJournal of Clinical Microbiology, 2001
- Comparative Performance of Three Viral Load Assays on Human Immunodeficiency Virus Type 1 (HIV-1) Isolates Representing Group M (Subtypes A to G) and Group O: LCx HIV RNA Quantitative, AMPLICOR HIV-1 MONITOR Version 1.5, and Quantiplex HIV-1 RNA Version 3.0Journal of Clinical Microbiology, 2001