An extensive log‐file analysis of step‐and‐shoot intensity modulated radiation therapy segment delivery errors
- 26 May 2004
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 31 (6) , 1593-1602
- https://doi.org/10.1118/1.1751011
Abstract
We present a study to evaluate the monitor unit (MU), dosimetric, and leaf-motion errors found in the delivery of 91 step-and-shoot IMRTtreatment plans performed at three nominal dose rates using a dual modality high energy Linac (Varian 2100 C/D, Varian MedicalSystems Inc., Palo Alto, CA) equipped with a 120-leaf multileaf collimator(MLC). The analysis was performed by studying log files generated by the MLC controller system. Recent studies by our group have validated that the automatically generated MLC log files accurately record the actual systemdelivery. A total of 635 beams were delivered at three nominal dose rates: 100, 300, and 600 MU/min. The log files were manually retrieved and analysis software was developed to extract the recorded MU delivery and leaf positions for each segment. Our analysis revealed that the magnitude of segment MU errors were independent of the planned segment MUs. Segment MU errors were found to increase with dose rate having maximum errors per segment of ±1.8 MU at 600 MU/min, ±0.8 MU at 300 MU/min, and ±0.5 MU at 100 MU/min. The total absolute MU error in each plan was observed to increase with the number of plan segments, with the trend increasing more rapidly for higher dose rates. Three dimensional dose distributions were recomputed based on the observed segment MU errors for three plans with large cumulative absolute MU errors. Comparison with the original treatment plans indicated no clinically significant consequences due to these errors. In addition, approximately 80% of the total segment deliveries reported at least one collimator leaf moving at least 1 mm (projected at isocenter) during segment delivery. Such errors occur near the end of segment delivery and have been previously observed by our group using a fast video-based electronic portal imaging device. At 600 MU/min, between 5% and 23% of the plan MUs were delivered during leaf motion that had exceeded a 1 mm position tolerance. These leaf motion errors were not included in the treatment plan recalculations performed in this study.Keywords
This publication has 13 references indexed in Scilit:
- Verification of step-and-shoot IMRT delivery using a fast video-based electronic portal imaging deviceMedical Physics, 2004
- Validation of dynamic MLC-controller log files using a two-dimensional diode arrayMedical Physics, 2003
- Incorporation of realistic delivery limitations into dynamic MLC treatment deliveryMedical Physics, 2002
- Verification of dynamic and segmental IMRT delivery by dynamic log file analysisJournal of Applied Clinical Medical Physics, 2002
- Communication and sampling rate limitations in IMRT delivery with a dynamic multileaf collimator systemMedical Physics, 2002
- Comprehensive quality assurance for the delivery of intensity modulated radiotherapy with a multileaf collimator used in the dynamic modeMedical Physics, 2001
- Intensity-modulated radiotherapy: current status and issues of interestInternational Journal of Radiation Oncology*Biology*Physics, 2001
- The overshoot phenomenon in step-and-shoot IMRT deliveryJournal of Applied Clinical Medical Physics, 2001
- Characterization of a commercial multileaf collimator used for intensity modulated radiation therapyMedical Physics, 2001
- Leaf position error during conformal dynamic arc and intensity modulated arc treatmentsMedical Physics, 2001