A New Mock Circulatory Loop and Its Application to the Study of Chemical Additive and Aortic Pressure Effects on Hemolysis in the Penn State Electric Ventricular Assist Device
- 1 May 1994
- journal article
- Published by Wiley in Artificial Organs
- Vol. 18 (5) , 397-407
- https://doi.org/10.1111/j.1525-1594.1994.tb02222.x
Abstract
A new mock circulatory loop was developed for hemolysis studies associated with the Penn State electric ventricular assist device (EVAD). This flow loop has several advantages over previously designed loops. It is small enough to accommodate experiments in which only single units of blood are available, it is made out of biocompatible materials, it incorporates good geometry, and it provides normal physiological pressures and flows to both the aortic outlet and the venous inlet of the pumping device. Experiments with reduced aortic pressure but normal cardiac output showed that hemolysis in a loop with normal aortic blood pressure was significantly higher than that in a loop with lowered aortic pressure, thereby illustrating the importance of maintaining loop pressures as close as possible to those found in vivo. This data also imply that blood traveling through the left ventricle in an artificial heart may be subject to higher hemolysis rates than that traversing the right ventricle. Another set of experiments to determine the effects of 4 hemolysis or drag-reducing agents (Pluronic F-68, Dextran-40, Polyox WSR-301, and Praestol 2273TR) on blood trauma due to the EVAD and associated valves was performed. Results indicated that none of the additives significantly reduced hemolysis under the conditions found in the mock loop. Finally, a compilation of data gathered in these experiments showed that the index of hemolysis (IH) is dependent on hematocrit (HCT), which suggests that another parameter, IH/HCT, may be more suited to the quantification of hemolysis.Keywords
This publication has 11 references indexed in Scilit:
- The Effect of Valve Type and Drive Line dP/dt on Hemolysis in the Pneumatic Ventricular Assist DeviceArtificial Organs, 1994
- Hematologic and Hemostatic Complications Associated with Long‐Term Use of Total Artificial Heart: Clinical and Experimental ObservationsArtificial Organs, 1992
- Effect of Hematocrit on Wall Shear Rate in Oscillatory Flow: Do the Elastic Properties of Blood Play a Role?Biorheology, 1991
- Media additives for protecting freely suspended animal cells against agitation and aeration damageTrends in Biotechnology, 1991
- An In Vitro Evaluation of an Artificial HeartAnnual Northeast Bioengineering Conference, 1991
- Hemodynamic Effects of a Poly(Ethylene Oxide) Drag-Reducing Polymer, Poly ox WSR N-60K, in the Open-Chest RatJournal of Cardiovascular Pharmacology, 1989
- Dextran protection of erythrocytes from low‐pH‐induced hemolysisFEBS Letters, 1989
- Decreased poststenotic flow disturbance during drag reduction by polyacrylamide infusion without increased aortic blood flowMicrovascular Research, 1989
- Reduction of hemolytic blood damage with dextranAIChE Journal, 1987
- Adjunctive Use of a Surface-Active Agent in Extracorporeal CirculationCirculation, 1966