Dynamic flat panel detector versus image intensifier in cardiac imaging: dose and image quality
- 23 November 2005
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 50 (23) , 5731-5742
- https://doi.org/10.1088/0031-9155/50/23/022
Abstract
The practical aspects of the dosimetric and imaging performance of a digital x-ray system for cardiology procedures were evaluated. The system was configured with an image intensifier (II) and later upgraded to a dynamic flat panel detector (FD). Entrance surface air kerma (ESAK) to phantoms of 16, 20, 24 and 28 cm of polymethyl methacrylate (PMMA) and the image quality of a test object were measured. Images were evaluated directly on the monitor and with numerical methods (noise and signal-to-noise ratio). Information contained in the DICOM header for dosimetry audit purposes was also tested. ESAK values per frame (or kerma rate) for the most commonly used cine and fluoroscopy modes for different PMMA thicknesses and for field sizes of 17 and 23 cm for II, and 20 and 25 cm for FD, produced similar results in the evaluated system with both technologies, ranging between 19 and 589 microGy/frame (cine) and 5 and 95 mGy min(-1) (fluoroscopy). Image quality for these dose settings was better for the FD version. The 'study dosimetric report' is comprehensive, and its numerical content is sufficiently accurate. There is potential in the future to set those systems with dynamic FD to lower doses than are possible in the current II versions, especially for digital cine runs, or to benefit from improved image quality.Keywords
This publication has 11 references indexed in Scilit:
- Dose performance evaluation of a charge coupled device and a flat-panel digital fluoroscopy system recently installed in an interventional cardiology laboratoryRadiation Protection Dosimetry, 2004
- Comparison of a conventional and a flat-panel digital system in interventional cardiology proceduresThe British Journal of Radiology, 2004
- Signal detectability in digital radiography: Spatial domain figures of meritMedical Physics, 2003
- Flachbilddetektoren in der R ntgendiagnostikDer Radiologe, 2003
- Training in radiological protection for interventionalists. Initial Spanish experienceThe British Journal of Radiology, 2003
- Patient Dose Related to the Complexity of Interventional Cardiology ProceduresRadiation Protection Dosimetry, 2001
- Introduction to constancy check protocols in fluoroscopic systems.Radiation Protection Dosimetry, 2001
- Clinical and technical determinants of the complexity of percutaneous transluminal coronary angioplasty procedures: Analysis in relation to radiation exposure parametersCatheterization and Cardiovascular Interventions, 2000
- Digital Radiography with a Large-Area, Amorphous-Silicon, Flat-Panel X-Ray Detector SystemInvestigative Radiology, 2000
- The optimization of signal detectability in digital fluoroscopyMedical Physics, 1982