Performance Study of Whole-Body, Multislice Positron Computed Tomograph POSITOLOGICA-II
- 1 February 1983
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 30 (1) , 734-738
- https://doi.org/10.1109/tns.1983.4332367
Abstract
A 3-detector ring, 5-slice whole-body positron CT has been developed and is being tested. The PCT, POSITOLOGICA-II, has a total of 480 BGO's (160/ring), and employs continuous rotation scan (0.5rps). By using a 15mm wide BGO, a 9.2mm FWHM spatial resolution for reconstructed image is obtained at the center of FOV. Measured phantom diameter dependence of the true count rate shows good agreement with theoretically anticipated characteristics, including maximum sensitivity at around 30cm dia. Sensitivities for 20cm dia. phantom are 28 and 38 kcps/μCi/ml for in-plane and cross-plane, respectively, including scattered coincidences. Relative system detection efficiency measured with a line source at FOV center is 96% at 15ns time window (90% at 12ns), basing on 100% efficiency in 20-24ns.Keywords
This publication has 7 references indexed in Scilit:
- A PROCEDURE FOR CALIBRATING AND CORRECTING DATA TO ACHIEVE ACCURATE QUANTITATIVE VALUES IN POSITRON EMISSION TOMOGRAPHYJournal of Computer Assisted Tomography, 1982
- Quad-Detector Arrangement and Sampling Characteristics in Rotary Positron Tomography: POSITOLOGICA IIIEEE Transactions on Medical Imaging, 1982
- Analytical Study of the Performance of a Multilayer Positron Computed Tomography ScannerJournal of Computer Assisted Tomography, 1982
- A quad BGO detector and its timing and positioning discrimination for positron computed tomographyNuclear Instruments and Methods in Physics Research, 1982
- Evaluation of Slice Shield Collimators for Multi-Layer Positron Emission Computed TomographsIEEE Transactions on Nuclear Science, 1982
- Design Consideration for a Continuously Rotating Positron Computed TomographIEEE Transactions on Nuclear Science, 1982
- Evaluating the Performance of Multiplane Positron Tomographs Designed for Brain ImagingIEEE Transactions on Nuclear Science, 1982