Impact of multileaf collimator leaf positioning accuracy on intensity modulation radiation therapy quality assurance ion chamber measurements
- 27 April 2005
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 32 (5) , 1440-1445
- https://doi.org/10.1118/1.1901843
Abstract
Quality assurance (QA) procedures for intensity modulation radiation therapy (IMRT) usually involve an ion chamber measurement in a phantom using the beam configuration of the actual treatment plan. In our QA procedures it was observed that the degree of agreement between the measurement and the calculation could vary from plan to plan, from linac to linac, as well as over time, with a discrepancy up to 8%. In this paper we examine one aspect of the process which can contribute to such poor reproducibility, namely, the leaf end position accuracy. A series of measurements was designed to irradiate an ion chamber using small beam segments where one multileaf collimator (MLC) edge covers half of the chamber. It was shown that the reproducibility varied up to 13%, which provides a possible explanation for the observed discrepancies above. A useful tool was also developed to measure ionization signals from individual segments of an IMRT sequence. In addition, an understanding of the leaf end position variations offers some insight into the overall quality of an IMRT dose distribution.Keywords
This publication has 14 references indexed in Scilit:
- The behavior of several microionization chambers in small intensity modulated radiotherapy fieldsMedical Physics, 2004
- Ionization chamber‐based reference dosimetry of intensity modulated radiation beamsMedical Physics, 2004
- An EGSnrc Monte Carlo study of the microionization chamber for reference dosimetry of narrow irregular IMRT beamletsMedical Physics, 2004
- An extensive log‐file analysis of step‐and‐shoot intensity modulated radiation therapy segment delivery errorsMedical Physics, 2004
- Verification of step-and-shoot IMRT delivery using a fast video-based electronic portal imaging deviceMedical Physics, 2004
- Guidance document on delivery, treatment planning, and clinical implementation of IMRT: Report of the IMRT subcommittee of the AAPM radiation therapy committeeMedical Physics, 2003
- Comparison of ionization chambers of various volumes for IMRT absolute dose verificationMedical Physics, 2003
- A method for geometrical verification of dynamic intensity modulated radiotherapy using a scanning electronic portal imaging deviceMedical Physics, 2002
- Recording accelerator monitor units during electronic portal imaging: application to collimator position verification during IMRTPhysics in Medicine & Biology, 2001
- Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapyMedical Physics, 1998