Abstract
The authors investigated the importance of 2D versus 3D compensation methods in SPECT. The compensation methods included in the study addressed two important degrading factors, namely attenuating and collimator-detector response in SPECT. They can be divided into two general categories. The conventional methods are based on the filtered backprojection algorithm, the Chang algorithm for attenuation compensation and the Metz filter for detector response compensation. These methods, which were computationally efficient, could only achieve approximate compensation due to the assumptions made. The quantitative compensation methods provide accurate compensation by modelling the degrading effects at the expense of large computational requirements. Both types of compensation methods were implemented in 2D and 3D reconstructions. The 2D and 3D reconstruction/compensation methods were evaluated using data from simulation of brain and heart, and patient thallium SPECT studies. The authors' results demonstrate the importance of compensation methods in improving the quality and quantitative accuracy of SPECT images and the relative effectiveness of the different 2D and 3D reconstruction/compensation methods. It is concluded that 3D implementation of the quantitative compensation methods provides the best SPECT image in terms of quantitative accuracy, spatial resolution, and noise at a cost of high computational requirements.Keywords
This publication has 37 references indexed in Scilit:
- Noise properties of filtered-backprojection and ML-EM reconstructed emission tomographic imagesIEEE Transactions on Nuclear Science, 1993
- A fast projector-backprojector pair modeling the asymmetric, spatially varying scatter response function for scatter compensation in SPECT imagingIEEE Transactions on Nuclear Science, 1993
- Quantitative SPECT brain imaging: effects of attenuation and detector responseIEEE Transactions on Nuclear Science, 1993
- Reconstruction methods for quantitative brain SPECTIEEE Transactions on Nuclear Science, 1993
- Comparison between ML-EM and WLS-CG algorithms for SPECT image reconstructionIEEE Transactions on Nuclear Science, 1991
- Three-dimensional iterative reconstruction algorithms with attenuation and geometric point response correctionIEEE Transactions on Nuclear Science, 1991
- Practical tradeoffs between noise, quantitation, and number of iterations for maximum likelihood-based reconstructionsIEEE Transactions on Medical Imaging, 1991
- Parameterization of the scatter response function in SPECT imaging using Monte Carlo simulationIEEE Transactions on Nuclear Science, 1990
- Improving the efficiency of emission tomography simulations using variance reduction techniquesIEEE Transactions on Nuclear Science, 1990
- Energy and spatial distribution of multiple order Compton scatter in SPECT: a Monte Carlo investigationPhysics in Medicine & Biology, 1984