Practical tradeoffs between noise, quantitation, and number of iterations for maximum likelihood-based reconstructions
- 1 January 1991
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Medical Imaging
- Vol. 10 (4) , 563-571
- https://doi.org/10.1109/42.108591
Abstract
Emission computerised tomography images reconstructed using a maximum likelihood-expectation maximization (ML)-based method with different reconstruction kernels and 1-200 iterations are compared to images reconstructed using filtered backprojection (FBP). ML-based reconstructions using a single pixel (SP) kernel with or without a sieve filter show no quantitative advantage over FBP except in the background where a reduction of noise is possible if the number of iterations is kept small (<50). ML-based reconstructions using a Gaussian kernel with a multipixel full-width-at-half-maximum (FWHM) and a large number of iterations (200) require a sieve filtering step to reduce the noise and contrast overshoot in the final images. These images have some small quantitative advantages over FBP depending on the structures being imaged. It is demonstrated that a feasibility stopping criterion controls the noise in a reconstructed image, but is insensitive to quantitation errors, and that the use of an appropriate overrelaxation parameter can accelerate the convergence of the ML-based method during the iterative process without quantitative instabilities.Keywords
This publication has 17 references indexed in Scilit:
- Image improvements in positron-emission tomography due to measuring differential time-of-flight and using maximum-likelihood estimationIEEE Transactions on Nuclear Science, 1990
- Feasible images and practical stopping rules for iterative algorithms in emission tomographyIEEE Transactions on Medical Imaging, 1989
- Quantitation of PET data with the EM reconstruction techniqueIEEE Transactions on Nuclear Science, 1989
- MLE reconstruction of a brain phantom using a Monte Carlo transition matrix and a statistical stopping ruleIEEE Transactions on Nuclear Science, 1988
- Stopping Rule for the MLE Algorithm Based on Statistical Hypothesis TestingIEEE Transactions on Medical Imaging, 1987
- Noise and Edge Artifacts in Maximum-Likelihood Reconstructions for Emission TomographyIEEE Transactions on Medical Imaging, 1987
- A Fast Reconstruction Algorthm for Stationary Positron Emission Tomography Based on a Modified EM AlgorithmIEEE Transactions on Medical Imaging, 1987
- The Use of Sieves to Stabilize Images Produced with the EM Algorithm for Emission TomographyIEEE Transactions on Nuclear Science, 1985
- Maximum Likelihood Reconstruction for Emission TomographyIEEE Transactions on Medical Imaging, 1982
- Algorithms for Fast Back- and Re-Projection in Computed TomographyIEEE Transactions on Nuclear Science, 1981