Cross-calibration of GE/Lunar pencil and fan-beam dual energy densitometers—bone mineral density and body composition studies
- 24 July 2003
- journal article
- Published by Springer Nature in European Journal of Clinical Nutrition
- Vol. 57 (8) , 977-987
- https://doi.org/10.1038/sj.ejcn.1601633
Abstract
In vitro and in vivo comparisons of bone mineral density (BMD) and body composition between GE/Lunar pencil (DPXL) and fan-beam (PRODIGY) absorptiometers. Comparison of BMD, bone mineral content (BMC) and area of lumbar spine (L2-L4), femoral neck and total body. Total body composition compartments tissue (TBTissue), fat (TBF), lean tissue (TBLean) and %TBF were also compared. Centre for Bone and Body Composition Research, University of Leeds. PHANTOMS/SUBJECTS: A range of spine phantoms, a variable composition phantom (VCP) and total body phantom. A total of 72 subjects were included for the in vivo study. In vitro: A small significant underestimation of BMD by the Prodigy compared to the DPXL ranging from 0.7 to 2% (p<0.05-0.001) for the spine phantoms. The Prodigy underestimated the VCP %Fat. Although the Prodigy underestimated phantom TBBMD by 1.1+/-1.0%, TBBMC and area were reduced by 8.2+/-1.4 and 7.3+/-1.0%, respectively. The Prodigy overestimated TBTissue 1508 g (2.2%), TBLean 588 g (1.2%), TBF 919 g (4.8%) and %TBF (0.8%). In vivo: BMD cross-calibration was only required in the femoral neck, DPXL(BMD)=0.08+0.906*PRODIGY(BMD). The Prodigy had higher estimates for TBTissue 1360 g (2.3%), TBLean 840 g (2.0%), TBF 519 g (3.4%), TBBMC 32.8 g (1.3%) and %TBF (0.3%). Cross-calibration equations were required for TBTissue(DPXL)=-1158+0.997*TBTissue(PRODIGY) and TBBMC(DPXL)= 89.7+0.949*TBBMC(PRODIGY). Small differences between the two absorptiometers for both BMD and body composition can be made compatible by use of cross-calibration equations and factors. The discrepancy in body composition compartments requires further research.Keywords
This publication has 29 references indexed in Scilit:
- Evaluation of a new body composition phantom for quality control and cross-calibration of DXA devicesJournal of Applied Physiology, 2000
- Bone Mineral and Body Composition Measurements: Cross-Calibration of Pencil-Beam and Fan-Beam Dual-Energy X-Ray AbsorptiometersJournal of Bone and Mineral Research, 1998
- Comparison of two Hologic DXA systems (QDR 1000 and QDR 4500/A).The British Journal of Radiology, 1997
- A multi-center comparison of dual energy X-ray absorptiometers: In vivo and in vitro soft tissue measurementEuropean Journal of Clinical Nutrition, 1997
- Cross-calibration of a fan-beam X-ray densitometer with a pencil-beam systemThe British Journal of Radiology, 1996
- Cross calibration of QDR-2000 and QDR-1000 dual-energy X-ray densitometers for bone mineral and soft-tissue measurementsBone, 1995
- Switching from DXA pencil-beam to fan-beam. I: Studies in vitro at four centersBone, 1994
- Switching from DXA pencil-beam to fan-beam. II: Studies in vivoBone, 1994
- Dual X-ray absorptiometry: a comparison between fan beam and pencil beam scansThe British Journal of Radiology, 1993
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986