Correction of eddy‐current distortions in diffusion tensor images using the known directions and strengths of diffusion gradients
Open Access
- 5 October 2006
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 24 (5) , 1188-1193
- https://doi.org/10.1002/jmri.20727
Abstract
Purpose: To correct eddy‐current artifacts in diffusion tensor (DT) images without the need to obtain auxiliary scans for the sole purpose of correction.Materials and Methods: DT images are susceptible to distortions caused by eddy currents induced by large diffusion gradients. We propose a new postacquisition correction algorithm that does not require any auxiliary reference scans. It also avoids the problematic procedure of cross‐correlating images with significantly different contrasts. A linear model is used to describe the dependence of distortion parameters (translation, scaling, and shear) on the diffusion gradients. The model is solved numerically to provide an individual correction for every diffusion‐weighted (DW) image.Results: The assumptions of the linear model were successfully verified in a series of experiments on a silicon oil phantom. The correction obtained for this phantom was compared with correction obtained by a previously published method. The algorithm was then shown to markedly reduce eddy‐current distortions in DT images from human subjects.Conclusion: The proposed algorithm can accurately correct eddy‐current artifacts in DT images. Its principal advantages are that only images with comparable signals and contrasts are cross‐correlated, and no additional scans are required. J. Magn. Reson. Imaging 2006.Keywords
This publication has 17 references indexed in Scilit:
- k‐space correction of eddy‐current‐induced distortions in diffusion‐weighted echo‐planar imagingMagnetic Resonance in Medicine, 2005
- Eddy current correction in diffusion‐weighted imaging using pairs of images acquired with opposite diffusion gradient polarityMagnetic Resonance in Medicine, 2003
- Reduction of eddy‐current‐induced distortion in diffusion MRI using a twice‐refocused spin echoMagnetic Resonance in Medicine, 2002
- A Model-Based Method for Retrospective Correction of Geometric Distortions in Diffusion-Weighted EPINeuroImage, 2002
- On the use of the FLAIR technique to improve the correction of eddy current induced artefacts in MR diffusion tensor imagingMagnetic Resonance Imaging, 2001
- On the use of water phantom images to calibrate and correct eddy current induced artefacts in MR diffusion tensor imagingMagnetic Resonance Imaging, 2000
- Mapping eddy current induced fields for the correction of diffusion-weighted echo planar imagesPublished by Elsevier ,1999
- Correction for distortion of echo‐planar images used to calculate the apparent diffusion coefficientMagnetic Resonance in Medicine, 1996
- Self-Compensating Pulsed Magnetic-Field Gradients for Short Recovery TimesJournal of Magnetic Resonance, Series A, 1994
- Estimation of the Effective Self-Diffusion Tensor from the NMR Spin EchoJournal of Magnetic Resonance, Series B, 1994