A Subcutaneous Lead Array for Implantable Cardioverter Defibrillators
- 1 July 1993
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 16 (7) , 1429-1443
- https://doi.org/10.1111/j.1540-8159.1993.tb01739.x
Abstract
Two patients received an implantable cardioverter defibrillator with the combination of a transvenous lead and a subcutaneous lead array with three branches. This approach allowed us to find low defibrillation threshoids in both patients (≤10 and ≤ 15 joules [J], respectively], which was impossible with a transvenous catheter. In a third patient, a crinkled subcutaneous patch was replaced by an array. The defibrillation threshoid with the array was ≤ 20 J, as opposed to ≤24 J with the patch. No surgical problems occurred. The subcutaneous array is a technical improvement for tiie therapy with impiantable defibrillators, when a single catheter system is not sufficient to ensure a safety margin for defibrillation, or when surgicai or postsurgical problems occur with a subcutaneous patch.Keywords
This publication has 7 references indexed in Scilit:
- Prospective comparison of biphasic and monophasic shocks for implantable cardioverter-defibrillators using endocardial leadsThe American Journal of Cardiology, 1992
- Assessment of the Defibrillation Threshold and Appropriate Sensing during Cardioverter Defibrillator ImplantationPacing and Clinical Electrophysiology, 1992
- Ventricular and Atrial Defibrillation Using New Transvenous Tripolar and Bipolar Leads with 5 French Electrodes and 8 French Subcutaneous CathetersPacing and Clinical Electrophysiology, 1991
- Patients with a high defibrillation threshold: Clinical characteristics, management, and outcomeAmerican Heart Journal, 1991
- Current Application of Implantable Cardioverter-Defibrillators in the Management of Malignant Ventricular TachyarrhythmiasCardiology, 1990
- Initial clinical experience with endocardial defibrillation using an implantable cardioverter/defibrillator with a triple-electrode systemArchives of internal medicine (1960), 1989
- Measurement of cardioversion/defibrillation thresholds in man by a truncated exponential waveform and an apical patch-superior vena caval spring electrode configuration.Circulation, 1984