Quality assurance for radiofrequency regional hyperthermia
- 1 January 1997
- journal article
- research article
- Published by Taylor & Francis in International Journal of Hyperthermia
- Vol. 13 (2) , 169-185
- https://doi.org/10.3109/02656739709012381
Abstract
Today most treatments with regional hyperthermia are applied using radiofrequency systems with 'focus' steering by amplitude and phase control. This paper deals with quality assurance procedures developed to ensure controlled and safe treatments in such systems. Our results show how the deviations between requested and observed phase and amplitude vary with frequency, and how these deviations depend on both the geometry of the object (phantom) inside the system and the power level applied. The results also indicate that the investigated systems' internal quality assurance procedures were inadequate and that additional procedures should be applied. Since the system parameters depend on patient and treatment specific conditions it is concluded that there is a need for QA measurements before or during treatment. This paper deals specifically with the commercial BSD-2000 system from BSD Medical Corp. in Salt Lake City, Utah, as installed in Bergen, but the procedure outlined can be applied to other phase and amplitude-controlled RF-RHT systems with only minimal adjustments.Keywords
This publication has 12 references indexed in Scilit:
- Verification of a hyperthermia model method using MR thermometryInternational Journal of Hyperthermia, 1995
- Deep Heating Using a Movable Applicator Phased Array Hyperthermia System: A preclinical feasibility studyActa Oncologica, 1994
- Multi-applicator hyperthermia system description using scattering parametersInternational Journal of Hyperthermia, 1994
- Letter to the EditorsInternational Journal of Hyperthermia, 1992
- Phantom studies and preliminary clinical experience with the BSD 2000International Journal of Hyperthermia, 1991
- Evaluation of the Sigma 60 applicator for regional hyperthermia in terms of scattering parametersInternational Journal of Hyperthermia, 1991
- Visualization by a matrix of light-emitting diodes of interference effects from a radiative four-applicator hyperthermia systemInternational Journal of Hyperthermia, 1991
- RTOG quality assurance guidelines for clinical trials using hyperthermiaInternational Journal of Radiation Oncology*Biology*Physics, 1990
- Quality assurance guidelines for ESHO protocolsInternational Journal of Hyperthermia, 1989
- Quality-assurance reports: CDRH RF phantom for hyperthermia systems evaluationsInternational Journal of Hyperthermia, 1988