Influence of Internal Current and Pacing Current on Pacemaker Longevity
- 1 January 1994
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 17 (1) , 13-16
- https://doi.org/10.1111/j.1540-8159.1994.tb01344.x
Abstract
The effects of lower pulse amplitude on battery current and pacemaker longevity were studied comparing the new, small-sized VVI pacemaker, Minix 8341, with the former model, Pasys 8329. Battery current was telemetrically measured at 0.8, 1.6, 2.5, and 5.0 V pulse amplitude and 0.05, 0.25, 0.5, and 1.0 msec pulse duration. Internal current was assumed to be equal to the battery current at 0.8 V and 0.05 msec. Pacing current was calculated subtracting internal current from battery current. The Minix pacemaker had a significantly lower battery current because of a lower internal current (Minix: 4.1 +/- 0.1 microA; Pasys: 16.1 +/- 0.1 microA); pacing current of both units was similar. At 0.5 msec pulse duration, the programming from 5.0-2.5 V pulse amplitude resulted in a greater relative reduction of battery current in the newer pacemaker (51% vs 25%). Projected longevity of each pacemaker was 7.9 years at 5.0 V and 0.5 msec. The programming from 5.0-2.5 V extended the projected longevity by 2.3 years (Pasys) and by 7.1 years (Minix). The longevity was negligibly longer after programming to 1.6 V. extension of pacemaker longevity can be achieved with the programming to 2.5 V or less if the connected pacemakers need a low internal current for their circuitry.Keywords
This publication has 12 references indexed in Scilit:
- Performance of Implantable Cardiac Rhythm Management DevicesPacing and Clinical Electrophysiology, 1992
- The Safety and Efficacy of Chronic Ventricular Pacing at 1.6 Volts Using a Steroid Eluting LeadPacing and Clinical Electrophysiology, 1992
- Benefits of Smaller Electrode Surface Area (4 mm2) on Steroid-Eluting LeadsPacing and Clinical Electrophysiology, 1991
- Energy‐conserving programming of vvi pacemakers: A telemetry‐supported, long‐term, follow‐up studyClinical Cardiology, 1990
- Stimulation Characteristics of a Steroid-Eluting Electrode Compared with Three Conventional ElectrodesPacing and Clinical Electrophysiology, 1990
- Lower Chronic Stimulation Threshold in the Carbon-Tip than in the Platinum-Tip Endocardial Electrode: A Randomized StudyPacing and Clinical Electrophysiology, 1989
- Survival Rates after Pacemaker Implantation: A Study of Patients Paced for Sick Sinus Syndrome and Atrioventricuiar BlockPacing and Clinical Electrophysiology, 1989
- A Comparison of Platinized Grooved Electrode Performance with Ring‐Tip ElectrodesPacing and Clinical Electrophysiology, 1989
- Acute and Long‐Term Atrial and Ventricular Stimulation Thresholds with a Steroid‐Eluting ElectrodePacing and Clinical Electrophysiology, 1985
- The present status of programmable pacemakersProgress in Cardiovascular Diseases, 1981