True three-dimensional image reconstruction by nuclear magnetic resonance zeugmatography
- 1 September 1981
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 26 (5) , 851-856
- https://doi.org/10.1088/0031-9155/26/5/004
Abstract
Nuclear magnetic resonance (NMR) zeugmatographic imaging may become a safe and versatile alternative to medical imaging techniques that employ ionising and ultrasonic radiation. Most of the techniques that have been described for obtaining NMR images use single point, line, or plane scans to give a single slice, or reconstruct only a two-dimensional projection, and are relatively inefficient, complex, or difficult to scale up for use on the human body. There are a number of advantages to scanning simultaneously an approximately spherical volume to obtain a true three-dimensional image. A simple two-stage reconstruction method is described for obtaining such images efficiently with isotropic resolution, and examples are presented to demonstrate the validity and usefulness of this mode. The feasibility of high-resolution imaging on large objects is also discussed.Keywords
This publication has 9 references indexed in Scilit:
- Automatic correction of nuclear magnetic resonance zeugmatographic projectionsJournal of Physics E: Scientific Instruments, 1981
- A gradient control device for complete three-dimensional nuclear magnetic resonance zeugmatographic imagingJournal of Physics E: Scientific Instruments, 1980
- Zeugmatography by Reconstruction from ProjectionsIEEE Transactions on Nuclear Science, 1980
- The sensitivity of the zeugmatographic experiment involving human samplesJournal of Magnetic Resonance (1969), 1979
- Thin-section NMR imagingJournal of Magnetic Resonance (1969), 1978
- NMR images by the multiple sensitive point method: application to larger biological systemsPhysics in Medicine & Biology, 1977
- Line scan proton spin imaging in biological structures by NMRPhysics in Medicine & Biology, 1976
- Image formation by nuclear magnetic resonance: The sensitive-point methodJournal of Applied Physics, 1976
- Image Formation by Induced Local Interactions: Examples Employing Nuclear Magnetic ResonanceNature, 1973