Wear of ultra-high molecular weight polyethylene acetabular cups in a physiological hip joint simulator in the anatomical position using bovine serum as a lubricant
- 1 March 1997
- journal article
- research article
- Published by SAGE Publications in Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
- Vol. 211 (3) , 265-269
- https://doi.org/10.1243/0954411971534377
Abstract
The Leeds physiological anatomical (PA) hip joint simulator was developed to apply three axes of loading and a complex three-dimensional motion so that the forces and motions can reproduce exactly the walking cycles defined by Paul. This paper presents the results of a study using the Leeds PA hip joint simulator to determine the wear of 32 mm ultra-high molecular weight polyethylene (UHMWPE) acetabular cups against stainless steel and zirconia ceramic heads, using bovine serum as lubricant. These results have been compared with the results of a previous study that used water as the lubricant, which led to UHMWPE transfer film being formed on the stainless steel head. Comparisons are also made with clinical results and results from other simulators. The study indicates that it is preferable to use bovine serum in simulator studies. In addition, the results indicate that if the surface roughness of the metallic and femoral heads are similar, and they remain undamaged during the tests, the wear rates of the UHMWPE cups are likely to be similar.Keywords
This publication has 11 references indexed in Scilit:
- Mechanistic and Morphological Origins of Ultra-High Molecular Weight Polyethylene Wear Debris in Total Joint Replacement ProsthesesProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1996
- Preliminary study of the effect of aging following irradiation on the wear of ultrahigh-molecular-weight polyethyleneThe Journal of Arthroplasty, 1995
- Assessment of the Change in Volume of Acetabular Cups Using a Coordinate Measuring MachineProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1994
- Comparative study of the wear of UHMWPE with zirconia ceramic and stainless steel femoral heads in artificial hip jointsMedical Engineering & Physics, 1994
- In vivo wear of polyethylene acetabular componentsThe Journal of Bone and Joint Surgery. British volume, 1993
- Wear of the polyethylene acetabular cup: Metallic and ceramic heads compared in a hip simulatorActa Orthopaedica, 1993
- Effect of femoral head size on wear of the polyethylene acetabular component.Journal of Bone and Joint Surgery, 1990
- The Influence of Bone and Bone Cement Debris on Counterface Roughness in Sliding Wear Tests of Ultra-High Molecular Weight Polyethylene on Stainless SteelProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1990
- Wear of ion-implanted pure titanium against UHMWPEBiomaterials, 1989
- The Role of Cement in the Long Term Performance and Premature Failure of Charnley Low Friction ArthroplastiesEngineering in Medicine, 1986