Area Under the Real‐Time Contact Force Curve (Force–Time Integral) Predicts Radiofrequency Lesion Size in an In Vitro Contractile Model
Top Cited Papers
- 2 September 2010
- journal article
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 21 (9) , 1038-1043
- https://doi.org/10.1111/j.1540-8167.2010.01750.x
Abstract
Electrode tissue contact, radiofrequency (RF) power and duration are major determinants of RF lesion size. Since contact forces (CF) vary in the beating heart, we evaluated contact force-time integral (FTI) as a predictor of lesion size at constant RF power in a contractile bench model simulating the beating heart. An open-tip irrigated catheter was attached to a movable mount incorporating a dynamic force sensor allowing closed loop control to achieve desired force variations between the catheter tip and bovine skeletal muscle placed on a ground plate. RF energy (20 and 40 W for 60 seconds, 17 cc/min irrigation) was delivered during (1) constant contact (C) at 20 g, (2) variable contact (V) with a 20 g peak and 10 g nadir, and (3) intermittent contact (I) with a 20 g peak and 0 g nadir with loss of contact. V and I protocols were performed at 50 and 100 catheter movements/min and 2 systole:diastole time ratios (50:50 and 30:70). The area under the CF curve was calculated as the FTI. Measured FTI was highest in C, intermediate during V and lowest during I and correlated linearly with lesion volume (P < 0.0001 for 20 and 40 W). Lesion volume was highest in group C, intermediate in V and lowest in group I (P < 0.05 for C vs V, V vs I, and C vs I). Lesion size correlates linearly with measured contact FTI. Constant contact produces the largest and intermittent contact the smallest lesions despite constant RF power and identical peak contact forces.Keywords
This publication has 6 references indexed in Scilit:
- Safety Profiles and Lesion Size of Different Radiofrequency Ablation Technologies: A Comparison of Large Tip, Open and Closed Irrigation CathetersJournal of Cardiovascular Electrophysiology, 2009
- Novel Contact Force Sensor Incorporated in Irrigated Radiofrequency Ablation Catheter Predicts Lesion Size and Incidence of Steam Pop and ThrombusCirculation: Arrhythmia and Electrophysiology, 2008
- AB36-6: Evaluation of a new catheter sensor for real-time measurement of tissue contactHeart Rhythm, 2006
- The Effects of Electrode‐Tissue Contact on Radiofrequency Lesion GenerationPacing and Clinical Electrophysiology, 1997
- Determinants of Lesion Size During Radiofrequency Catheter Ablation: The Role of Electrode‐Tissue Contact Pressure and Duration of Energy DeliveryJournal of Cardiovascular Electrophysiology, 1991
- Tissue Heating During Radiofrequency Catheter Ablation: A Thermodynamic Model and Observations in Isolated Perfused and Superfused Canine Right Ventricular Free WallPacing and Clinical Electrophysiology, 1989