Scatter correction in SPECT using non-uniform attenuation data
- 1 October 1988
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 33 (10) , 1129-1140
- https://doi.org/10.1088/0031-9155/33/10/003
Abstract
Quantitative assessment of activity levels with SPECT is difficult because of attenuation and scattering of gamma rays within the object. To study the effect of attenuation and scatter on SPECT quantitation, phantom studies were performed with non-uniform attenuation. Simulated transmission CT data provided information about the distribution of attenuation coefficients within the source. Attenuation correction was performed by an iterative reprojection technique. Scatter correction was done by convolution of the attenuation-corrected image and an appropriate filter. The filter characteristics depended on the attenuation and activity measurement at each pixel. The scatter correction could compensate completely for the 28% scatter component from a line source, and the 61% component from a thick, extended source. Accuracy of regional activity ratios and the linearity of the relationship between true radioactivity and the SPECT measurement were both significantly improved by these corrections. The present method is expected to be valuable for the quantitative assessment of regional activity.This publication has 12 references indexed in Scilit:
- PERFORMANCE EVALUATION OF SPRINT, A SINGLE PHOTON RING TOMOGRAPH FOR BRAIN IMAGING1984
- IMPROVED SPECT QUANTIFICATION USING COMPENSATION FOR SCATTERED PHOTONS1984
- Analysis of SPECT including Scatter and Attenuation Using Sophisticated Monte Carlo Modeling MethodsIEEE Transactions on Nuclear Science, 1982
- Attenuation Correction in Gamma Emission Computed TomographyJournal of Computer Assisted Tomography, 1981
- DETERMINATION OF THE PLEURAL EDGE BY GAMMA-RAY TRANSMISSION COMPUTED-TOMOGRAPHY1981
- An Integral-Transport Method for Compton-Scatter Correction in Emission Computed TomographyIEEE Transactions on Nuclear Science, 1980
- Quantitation in Positron Emission Computed TomographyJournal of Computer Assisted Tomography, 1979
- A New Approach to the Emission Computerized Tomography Problem: Simultaneous Calculation of Attenuation and Activity CoefficientsIEEE Transactions on Nuclear Science, 1979
- A Method for Attenuation Correction in Radionuclide Computed TomographyIEEE Transactions on Nuclear Science, 1978
- A Quantitative Theory of the Hounsfield Unit and Its Application to Dual Energy ScanningJournal of Computer Assisted Tomography, 1977