Active deep brain stimulation during MRI: A feasibility study
Open Access
- 26 January 2004
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 51 (2) , 380-388
- https://doi.org/10.1002/mrm.10699
Abstract
The goal of this study was to evaluate the feasibility of active deep brain stimulation (DBS) during the application of standard clinical sequences for functional MRI (fMRI) in phantom measurements. During active DBS, we investigated induced voltage, temperature at the electrode tips and lead, forces on the electrode and lead, consequences of defective leads and loose connections, proper operation of the neurostimulator, and image quality. Sequences for diffusion‐ and perfusion‐weighted imaging, fMRI, and morphologic MRI were used. The DBS electrode and lead were placed in a NaCl solution‐filled phantom. The results indicate that there are severe potential hazards for patients. Strong heating, high induced voltage, and even sparking at defects in the connecting cable could be observed. However, it was demonstrated that under certain conditions, safe MR examinations during active DBS are feasible. Certain safety precautions are recommended in this report. Magn Reson Med 51:380–388, 2004.Keywords
This publication has 12 references indexed in Scilit:
- Hazardous situation in the MR bore: induction in ECG leads causes fireEuropean Radiology, 2003
- Radiofrequency heating effects around resonant lengths of wire in MRIPhysics in Medicine & Biology, 2002
- Electrical and thermal behavior of non-ferrous noble metal electrodes exposed to MRI fieldsMagnetic Resonance Imaging, 2002
- Neurostimulation systems for deep brain stimulation: In vitro evaluation of magnetic resonance imaging–related heating at 1.5 teslaJournal of Magnetic Resonance Imaging, 2002
- Implantable spinal fusion stimulator: Assessment of MR safety and artifactsJournal of Magnetic Resonance Imaging, 2000
- Thalamic stimulation and functional magnetic resonance imaging: localization of cortical and subcortical activation with implanted electrodesJournal of Neurosurgery, 1999
- Magnetic Resonance Imaging with Implanted Neurostimulators: An In Vitro and In Vivo StudyNeurosurgery, 1999
- Echo-Planar Imaging Image ArtifactsPublished by Springer Nature ,1998
- Deep brain stimulation: a review of basic research and clinical studiesPain, 1991
- Reversible Heat Lesions with Radiofrequency CurrentJournal of Neurosurgery, 1964