Analysis of the kinematics of different hip simulators used to study wear of candidate materials for the articulation of total hip arthroplasties
- 1 May 1998
- journal article
- research article
- Published by Wiley in Journal of Orthopaedic Research
- Vol. 16 (3) , 365-369
- https://doi.org/10.1002/jor.1100160313
Abstract
We developed an analytical technique to determine the paths traced by specific points on the femoral head against the acetabulum in the human hip joint during gait. The purpose of the study was to apply this technique to the mechanical hip simulators chosen to conduct wear tests on polymeric acetabular liners used in total hip replacements. These simulators differ from one another in the type of motion produced, apart from other variables such as type of lubricant and head position. Due to the variation in the kinematics between the machines, the paths traced by the points on the femoral head against the acetabular liner ranged from simple linear traces to figure-8 loops and quasi-elliptical paths during a single simulator cycle. The distances traveled by these points during the same period also varied appreciably among the different hip simulator designs. These results are important when combined with other studies that have shown that kinematics can play an important role in the outcome of in vitro wear experiments. The kinematic differences quantified in this study can partially explain the substantial differences in wear data reported from different simulator designs and also underscore the usefulness of the technique described in this study in judging the results from different hip simulator experiments.Keywords
This publication has 13 references indexed in Scilit:
- Orientation softening in the deformation and wear of ultra-high molecular weight polyethyleneWear, 1997
- Kinematics of the MATCO™ hip simulator and issues related to wear testing of metal-metal implantsProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1997
- Loci of movement of selected points on the femoral head during normal gaitThe Journal of Arthroplasty, 1996
- Mechanistic and Morphological Origins of Ultra-High Molecular Weight Polyethylene Wear Debris in Total Joint Replacement ProsthesesProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1996
- Adaptiive finite element modeling of long‐term polyethylene wear in total hip arthroplastyJournal of Orthopaedic Research, 1996
- A Five-Station Hip Joint Simulator for Wear Rate StudiesProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1992
- Wear of ion-implanted pure titanium against UHMWPEBiomaterials, 1989
- Design and Development of a Versatile Hip Joint Simulator and a Preliminary Assessment of Wear and Creep in Charnley Total Replacement Hip JointsEngineering in Medicine, 1988
- Wear behaviour of medical engineering materialsWear, 1975
- A simulator for load bearing jointsWear, 1972