Transient dystonia following magnetic resonance imaging in a patient with deep brain stimulation electrodes for the treatment of Parkinson disease
- 1 October 2003
- journal article
- case report
- Published by Journal of Neurosurgery Publishing Group (JNSPG) in Journal of Neurosurgery
- Vol. 99 (4) , 772-774
- https://doi.org/10.3171/jns.2003.99.4.0772
Abstract
Data from previous studies have shown that magnetic resonance (MR) imaging of the head can be performed safely in patients with deep brain stimulators. The authors report on a 73-year-old patient with bilaterally implanted deep brain electrodes for the treatment of Parkinson disease, who exhibited dystonic and partially ballistic movements of the left leg immediately after an MR imaging session. Such dystonic or ballistic movements had not been previously observed in this patient. In the following months, this focal movement disorder resolved completely. This case demonstrates the possible risks of MR imaging in patients with deep brain stimulators.Keywords
This publication has 12 references indexed in Scilit:
- Functional magnetic resonance imaging during deep brain stimulation: A pilot study in four patients with Parkinson's diseaseMovement Disorders, 2001
- Effect of bilateral subthalamic electrical stimulation in Parkinson’s diseaseSurgical Neurology, 2001
- Investigation of the factors responsible for burns during MRIJournal of Magnetic Resonance Imaging, 2001
- Do unilateral ablative lesions of the subthalamic nucleu in parkinsonian patients lead to hemiballism?Movement Disorders, 2001
- The subthalamic nucleus, hemiballismus and Parkinson's disease: reappraisal of a neurosurgical dogmaBrain, 2001
- Lesioning the Subthalamic Nucleus in the Treatment of Parkinson’s DiseaseStereotactic and Functional Neurosurgery, 2001
- Symptomatic paroxysmal hemidystonia due to a demyelinating subthalamic lesionEuropean Journal of Neurology, 2000
- Thalamic stimulation and functional magnetic resonance imaging: localization of cortical and subcortical activation with implanted electrodesJournal of Neurosurgery, 1999
- Magnetic Resonance Imaging with Implanted Neurostimulators: An In Vitro and In Vivo StudyNeurosurgery, 1999
- The Effect of Magnetic Resonance Imagers on Implanted NeurostimulatorsPacing and Clinical Electrophysiology, 1992