Portal dose image (PDI) prediction for dosimetric treatment verification in radiotherapy. I. An algorithm for open beams
- 2 June 1998
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 25 (6) , 830-840
- https://doi.org/10.1118/1.598293
Abstract
A method is presented for calculation of transmission functions for high energy photon beams through patients. These functions are being used in our clinic for prediction of portal dose images (PDIs) which are compared with PDIs measured with an electronic portal imaging device (EPID). The calculations are based on the planning CT-scan of the patient and on the irradiation geometry as determined in the treatment planning process. For each beam quality, the required input data for the algorithm for transmission prediction are derived from a limited number of measured beam data. The method has been tested for a PDI-plane at 160 cm from the focus, in agreement with the fixed focus-to-detector distance of our fluoroscopic EPIDs. For 6, 23 and 25 MV photon beams good agreement (approximately 1%) has been found between calculated and measured transmissions through anthropomorphic phantoms.Keywords
This publication has 28 references indexed in Scilit:
- High-precision prostate cancer irradiation by clinical application of an offline patient setup verification procedure, using portal imagingInternational Journal of Radiation Oncology*Biology*Physics, 1996
- Calculation of the absorbed dose distribution due to irregularly shaped photon beams using pencil beam kernels derived from basic beam dataPhysics in Medicine & Biology, 1996
- Portal dose measurement in radiotherapy using an electronic portal imaging device (EPID)Physics in Medicine & Biology, 1995
- Daily dosimetric quality control of the MM50 Racetrack Microtron using an electronic portal imaging deviceRadiotherapy and Oncology, 1995
- Calculation models for determining the absorbed dose in water phantoms in off-axis planes of rectangular fields of open and wedged photon beamsPhysics in Medicine & Biology, 1995
- The use of an electronic portal imaging device for exit dosimetry and quality control measurementsInternational Journal of Radiation Oncology*Biology*Physics, 1995
- Accuracy of patient positioning in mantle field irradiationRadiotherapy and Oncology, 1992
- Quality assurance in radiotherapy by in vivo dosimetry. 1. Entrance dose measurements, a reliable procedureRadiotherapy and Oncology, 1990
- Evaluation of the Clinical Use of TLDActa Radiologica: Therapy, Physics, Biology, 1976
- Scatter-air ratiosPhysics in Medicine & Biology, 1972