Physical aspects of dynamic stereotactic radiosurgery with very small photon beams (1.5 and 3 mm in diameter)
- 10 January 2003
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 30 (2) , 111-118
- https://doi.org/10.1118/1.1536290
Abstract
Stereotactic radiosurgery is often used for treating functional disorders. For some of these disorders, the size of the target can be on the order of a millimeter and the radiation dose required for treatment on the order of 80 Gy. The very small radiation field and high prescribed dose present a difficult challenge in beam calibration, dose distribution calculation, and dose delivery. In this work the dose distribution for dynamic stereotactic radiosurgery, carried out with 1.5 and 3 mm circular fields, was studied. A 10 MV beam from a Clinac-18 linac (Varian, Palo Alto, CA) was used as the radiation source. The BEAM/EGS4 Monte Carlo code was used to model the treatment head of the machine along with the small-field collimators. The models were validated with the EGSnrc code, first through a calculation of percent depth doses (PDD) and dose profiles in a water phantom for the two small stationary circular beams and then through a comparison of the calculated with measured PDD and profile data. The three-dimensional (3-D) dose distributions for the dynamic rotation with the two small radiosurgical fields were calculated in a spherical water phantom using a modified version of the fast XVMC Monte Carlo code and the validated models of the machine. The dose distributions in a horizontal plane at the isocenter of the linac were measured with low-speed radiographic film. The maximum sizes of the Monte Carlo-calculated 50% isodose surfaces in this horizontal plane were 2.3 mm for the 1.5 mm diameter beam and 3.8 mm for the 3 mm diameter beam. The maximum discrepancies between the 50% isodose surface on the film and the 50% Monte Carlo-calculated isodose surfaces were 0.3 mm for both the 1.5 and 3 mm beams. In addition, the displacement of the delivered dose distributions with respect to the laser-defined isocenter of the machine was studied. The results showed that dynamic radiosurgery with very small beams has a potential for clinical use.Keywords
This publication has 10 references indexed in Scilit:
- Treatment of postherpetic trigeminal neuralgia with the Leksell gamma knifeJournal of Neurosurgery, 2000
- On isocentre adjustment and quality control in linear accelerator based radiosurgery with circular collimators and room lasersPhysics in Medicine & Biology, 2000
- Accurate condensed history Monte Carlo simulation of electron transport. I. EGSnrc, the new EGS4 versionMedical Physics, 2000
- Setup verification in linac‐based radiosurgeryMedical Physics, 1999
- Fast Monte Carlo dose calculation for photon beams based on the VMC electron algorithmMedical Physics, 1999
- 3D electron dose calculation using a Voxel based Monte Carlo algorithm (VMC)Medical Physics, 1996
- BEAM: A Monte Carlo code to simulate radiotherapy treatment unitsMedical Physics, 1995
- A system for stereotactic radiosurgery with a linear acceleratorInternational Journal of Radiation Oncology*Biology*Physics, 1988
- Dynamic stereotactic radiosurgeryInternational Journal of Radiation Oncology*Biology*Physics, 1988
- Cerebral radiation surgery using moving field irradiation at a linear accelerator facilityInternational Journal of Radiation Oncology*Biology*Physics, 1985