An evaluation of cardiac output by five arterial pulse contour techniques during cardiac surgery
- 13 July 2007
- journal article
- research article
- Published by Wiley in Anaesthesia
- Vol. 62 (8) , 760-768
- https://doi.org/10.1111/j.1365-2044.2007.05135.x
Abstract
The bias, precision and tracking ability of five different pulse contour methods were evaluated by simultaneous comparison of cardiac output values from the conventional thermodilution technique (COtd). The five different pulse contour methods included in this study were: Wesseling's method (cZ); the Modelflow method; the LiDCO system; the PiCCO system and a recently developed Hemac method. We studied 24 cardiac surgery patients undergoing uncomplicated coronary artery bypass grafting. In each patient, the first series of COtd was used to calibrate the five pulse contour methods. In all, 199 series of measurements were accepted by all methods and included in the study. COtd ranged from 2.14 to 7.55 l.min(-1), with a mean of 4.81 l.min(-1). Bland-Altman analysis showed the following bias and limits of agreement: cZ, 0.23 and - 0.80 to 1.26 l.min(-1); Modelflow, 0.00 and - 0.74 to 0.74 l.min(-1); LiDCO, - 0.17 and - 1.55 to 1.20 l.min(-1); PiCCO, 0.14 and - 1.60 to 1.89 l.min(-1); and Hemac, 0.06 and - 0.81 to 0.93 l.min(-1). Changes in cardiac output larger than 0.5 l.min(-1) (10%) were correctly followed by the Modelflow and the Hemac method in 96% of cases. In this group of subjects, without congestive heart failure, with normal heart rhythm and reasonable peripheral circulation, the best results in absolute values as well as in tracking changes in cardiac output were measured using the Modelflow and Hemac pulse contour methods, based on non-linear three-element Windkessel models.Keywords
This publication has 32 references indexed in Scilit:
- Monitoring cardiac output using the femoral and radial arterial pressure waveformAnaesthesia, 2006
- Pulseco: a less-invasive method to monitor cardiac output from arterial pressure after cardiac surgeryThe Annals of Thoracic Surgery, 2002
- Continuous stroke volume monitoring by modelling flow from non-invasive measurement of arterial pressure in humans under orthostatic stressClinical Science, 1999
- Continuous Cardiac Output in Septic Shock by Simulating a Model of the Aortic Input ImpedanceAnesthesiology, 1999
- Continuous, Less Invasive, Hemodynamic Monitoring in Intensive Care After Cardiac SurgeryThe Thoracic and Cardiovascular Surgeon, 1998
- Lithium dilution cardiac output measurementCritical Care Medicine, 1997
- An adequate strategy for the thermodilution technique in patients during mechanical ventilationIntensive Care Medicine, 1990
- Translation of Otto frank's paper “Die Grundform des arteriellen Pulses” zeitschrift für biologie 37: 483–526 (1899)Journal of Molecular and Cellular Cardiology, 1990
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- The pressure dependent dynamic elasticity of 35 thoracic and 16 abdominal human aortas in vitro described by a five component modelJournal of Biomechanics, 1985