Design, optimisation, and evaluation of a sheath flow interface for automated capillary electrophoresis‐electrospray‐mass spectrometry
- 11 April 2005
- journal article
- research article
- Published by Wiley in Electrophoresis
- Vol. 26 (7-8) , 1366-1375
- https://doi.org/10.1002/elps.200410133
Abstract
A sheath‐flow capillary electrophoresis‐mass spectrometry (CE‐MS) system utilizing a fully integrated large‐bore stainless‐steel emitter electrode tapered at the end for micro‐ionspray operation has been developed and evaluated. A separation capillary with an outer diameter of up to 360 µm was inserted into the electrode thus forming a void volume of less than 15 nL between the capillary end and the electrospray ionisation (ESI) tip. The sheath liquid, usually methanol‐water (80:20) with 0.1% formic acid for positive ion mode or methanol for negative ion mode, was delivered at 0.5–1.0 µL/min. Unlike previously reported CE‐MS interfaces, the CE‐MS probe was incorporated directly onto an Applied Biosystems/MDS SCIEX orthogonal‐spray Turbo “V” ion source for ease of use and automatic operation. This integration enables fast and facile coupling and replacement of the separation capillary without interrupting the ion source configuration, and the sheath liquid supply. The reusable electrospray electrode was precisely fabricated and aligned with the length of the nebulizing gas tube for improved reproducibility. Automation was achieved through software control of both CE and tandem MS (MS/MS) for unattended batch sample analysis. The system was evaluated for attomole‐ to low femtomole‐level profiling of model peptides and protein mixtures, bisphosphates, as well as antiviral nucleosidic drugs in cellular extracts.Keywords
This publication has 43 references indexed in Scilit:
- A sheath‐flow nanospray interface for capillary electrophoresis/mass spectrometryRapid Communications in Mass Spectrometry, 2004
- Capillary electrophoresis‐mass spectrometry: Recent advances to the analysis of small achiral and chiral solutesElectrophoresis, 2004
- On‐line capillary electrophoresis‐mass spectrometry for the analysis of biomoleculesElectrophoresis, 2004
- Capillary Electrophoresis to Mass Spectrometry Interface Using a Porous JunctionAnalytical Chemistry, 2003
- A robust micro‐electrospray ionization technique for high‐throughput liquid chromatography/mass spectrometry proteomics using a sanded metal needle as an emitterRapid Communications in Mass Spectrometry, 2002
- On‐line nonaqueous capillary electrophoresis and electrospray mass spectrometry of tricyclic antidepressants and metabolic profiling of amitriptyline by Cunninghamella elegansElectrophoresis, 1998
- Enhanced sensitivity for sequence determination of major histocompatibility complex class I peptides by membrane preconcentration ‐ capillary electrophoresis ‐microspray ‐ tandem mass spectrometryElectrophoresis, 1998
- Capillary Zone Electrophoresis−Electrospray Mass Spectrometry at Submicroliter Flow Rates: Practical Considerations and Analytical PerformanceAnalytical Chemistry, 1997
- Capillary electrophoresis/mass spectrometry determination of inorganic ions using an ion spray‐sheath flow interfaceElectrophoresis, 1993
- Sample introduction and separation in capillary electrophoresis, and combination with mass spectrometric detectionTalanta, 1989