Interactions between cochlear implant electrode insertion depth and frequency-place mapping
- 1 March 2005
- journal article
- conference paper
- Published by Acoustical Society of America (ASA) in The Journal of the Acoustical Society of America
- Vol. 117 (3) , 1405-1416
- https://doi.org/10.1121/1.1856273
Abstract
While new electrode designs allow deeper insertion and wider coverage in the cochlea, there is still considerable variation in the insertion depth of the electrode array among cochlear implant users. The present study measures speech recognition as a function of insertion depth, varying from a deep insertion of 10 electrodes at 28.8 mm to a shallow insertion of a single electrode at 7.2 mm, in four Med-El Combi users. Short insertion depths were simulated by inactivating apical electrodes. Speech recognition increased with deeper insertion, reaching an asymptotic level at 21.6 or 26.4 mm depending on the frequency-place map used. Başkent and Shannon [J. Acoust. Soc. Am. 116, 3130–3140 (2004)] showed that speech recognition by implant users was best when the acoustic input frequency was matched onto the cochlear location that normally processes that frequency range, minimizing the spectral distortions in the map. However, if an electrode array is not fully inserted into the cochlea, a matched map will result in the loss of considerable low-frequency information. The results show a strong interaction between the optimal frequency-place mapping and electrode insertion depth. Consistent with previous studies, frequency-place matching produced better speech recognition than compressing the full speech range onto the electrode array for full insertion ranges (20 to 25 mm from the round window). For shallower insertions (16.8 and 19.2 mm) a mild amount of frequency-place compression was better than truncating the frequency range to match the basal cochlear location. These results show that patients with shallow electrode insertions might benefit from a map that assigns a narrower frequency range than patients with full insertions.
Keywords
This publication has 22 references indexed in Scilit:
- Frequency-place compression and expansion in cochlear implant listenersThe Journal of the Acoustical Society of America, 2004
- Speech recognition under conditions of frequency-place compression and expansionThe Journal of the Acoustical Society of America, 2003
- Simulations of tonotopically mapped speech processors for cochlear implant electrodes varying in insertion depthThe Journal of the Acoustical Society of America, 2003
- Speech recognition in noise as a function of the number of spectral channels: Comparison of acoustic hearing and cochlear implantsThe Journal of the Acoustical Society of America, 2001
- Effects of Electrode Location and Spacing on Phoneme Recognition with the Nucleus-22 Cochlear ImplantEar & Hearing, 1999
- Simulating the effect of cochlear-implant electrode insertion depth on speech understandingThe Journal of the Acoustical Society of America, 1997
- Speech Recognition as a Function of the Number of Electrodes Used in the SPEAK Cochlear Implant Speech ProcessorJournal of Speech, Language, and Hearing Research, 1997
- Cochlear Implant Place PsychophysicsAudiology and Neurotology, 1996
- Middle-ear development V: Development of umbo sensitivity in the gerbilAmerican Journal of Otolaryngology, 1993
- Factors Predicting Postoperative Sentence Scores in Postlinguistically Deaf Adult Cochlear Implant PatientsAnnals of Otology, Rhinology & Laryngology, 1992