Comparison of the effect of different reference data on Lunar DPX and Hologic QDR-1000 dual-energy X-ray absorptiometers
- 1 December 1992
- journal article
- research article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 65 (780) , 1124-1129
- https://doi.org/10.1259/0007-1285-65-780-1124
Abstract
We have investigated whether the Lunar DPX (software 3.4) and Hologic QDR-1000 dual-energy X-ray absorptiometers have comparable normal reference databases for the spine and femur of white UK and USA subjects. After conversion for systematic differences in absolute bone density values between the two systems, the reference databases were very similar for the spine in young subjects, but there were clear differences in the femur databases of young females and males of all ages. These differences were confirmed by comparing the percent age-matched and young values determined by the two systems for subjects scanned on both systems. Thus the diagnosis and management of a patient could differ, depending on the system used for the bone density measurements.Keywords
This publication has 11 references indexed in Scilit:
- Measurements of bone mineral density of the proximal femur by two commercially available dual energy X-ray absorptiometric systemsEuropean Journal of Nuclear Medicine and Molecular Imaging, 1992
- Comparative performance in vitro and in vivo of Lunar DPX and Hologic QDR-1000 dual energy X-ray absorptiometersThe British Journal of Radiology, 1991
- Calibration of dual-energy X-ray absorptiometry for bone densityJournal of Bone and Mineral Research, 1991
- Intersite comparison of the Hologic QDR-1000 dual energy X-ray bone densitometerThe British Journal of Radiology, 1991
- The interchangeability of radioisotope and X-ray based measurements of bone mineral densityThe British Journal of Radiology, 1991
- Interpretation of bone densitometry measurements: Disadvantages of a percentage scale and a discussion of some alternativesJournal of Bone and Mineral Research, 1990
- Dual-energy radiographic absorptiometry of the lumbar spine: clinical experience with two different systems.Radiology, 1990
- X-ray dual-photon absorptiometry: a new method for the measurement of bone densityThe British Journal of Radiology, 1989
- Performance evaluation of a dual-energy X-ray bone densitometerCalcified Tissue International, 1989
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986