Dosimetric characteristics of the InterSource103 palladium brachytherapy source
- 5 May 2000
- journal article
- Published by Wiley in Medical Physics
- Vol. 27 (5) , 1093-1100
- https://doi.org/10.1118/1.598991
Abstract
103Pd brachytherapy sources are being used for interstitial implants in tumor sites such as the prostate. Recently, the InterSource103 palladium source has been introduced, which has a design different from that of other sources presently on the market. Dosimetric characteristics (i.e., dose rate constant, radial dose function, and anisotropy function) of this source were experimentally and theoretically determined following the AAPM Task Group 43 (TG-43) recommendations and were related to the 1999 NIST calibration assigned to this source [Sk, 99std]. Measurements were performed in a solid water phantom using LiF thermoluminescent dosimeters. The measured data was compared with Monte Carlo simulations performed in solid water using the PTRAN code. The calculations were then performed in liquid water to obtain the dosimetric information for clinical applications as per TG-43 recommendation. The results indicated that the dose rate constant, lambda, of the InterSource103 palladium source was 0.664+/-5% cGy/h/U using TLD measurements and 0.660+/-3% cGy/h/U using Monte Carlo simulations in solid water. The calculated value in liquid water was found to be 0.696 +/- 3 % cGy/h/U. The radial dose function, g(r), of the new 103Pd source was measured at distances ranging from 0.5 to 10 cm using LiF TLD in solid water phantom material. The Monte Carlo simulations were performed at distances ranging from 0.1 to 10 cm from the source center in solid water and liquid water. The anisotropy function, F(r, theta), was measured at distances of 2, 3, 5, and 7 cm from the source center and calculated at distances of 0.5, 1, 2, 3, 5, and 7 cm from the source center. Complete dosimetric data are described in this paper. Per TG-43, the values reported in water should be used for clinical treatment planning systems.Keywords
This publication has 18 references indexed in Scilit:
- Dosimetric characterization of a new design 103palladium brachytherapy sourceMedical Physics, 1999
- Guidance to users of Nycomed Amersham and North American Scientific, Inc., I-125 Interstitial Sources: Dosimetry and calibration changes: Recommendations of the American Association of Physicists in Medicine Radiation Therapy Committee Ad Hoc SubcommitteeMedical Physics, 1999
- Anisotropy functions for and sourcesMedical Physics, 1998
- Dosimetric prerequisites for routine clinical use of new low energy photon interstitial brachytherapy sourcesMedical Physics, 1998
- Prostate Specific Antigen Based Disease Control Following Ultrasound Guided sup 125 Iodine Implantation for Stage T1/T2 Prostatic CarcinomaJournal of Urology, 1995
- Experimental validation of Monte Carlo dose calculations about a high-intensity Ir-192 source for pulsed dose-rate brachytherapyMedical Physics, 1995
- Permanent interstitial implantation using palladium-103: The new york medical college preliminary experienceInternational Journal of Radiation Oncology*Biology*Physics, 1992
- Comparison of measured and calculated dose rates in water near I‐125 and Ir‐192 seedsMedical Physics, 1991
- Photon energy dependence of the sensitivity of radiochromic film and comparison with silver halide film and LiF TLDs used for brachytherapy dosimetryMedical Physics, 1991
- Thermoluminescent dosimetry for 1 0 3 Pd seeds (model 200) in solid water phantomMedical Physics, 1991