Output ratios in a miniphantom for asymmetric fields shaped by a multileaf collimator
- 1 November 1997
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 42 (11) , 2305-2317
- https://doi.org/10.1088/0031-9155/42/11/021
Abstract
The integrated GE multileaf collimator (MLC) provides the ability to achieve `double' asymmetric fields: each of the 64 leaves allow an over-axis travel of 10 cm and the Y-jaws allow 20 cm. A formalism has recently been proposed by the authors to calculate the output ratio in a miniphantom for this type of MLC by the product of independent leaf and jaw correction factors. The original proposed formalism was restricted to regular or irregular fields including the collimator rotational axis. Introducing `reduced coordinates' for the correction factors in the present work this formalism is extended to asymmetric fields where central leaves or jaws overlap the collimator axis. The extended formalism is applied to asymmetric square, rectangular and irregular fields. For all fields checked at a given off-axis position, measured and calculated output ratios agree within 1% for 6, 18 and 25 MV photon beams. To relate output ratios normalized to off-axis points with output ratios on-axis, off-axis ratios are derived from film and miniphantom measurements. Both off-axis ratios agree to within 1% for 6 and 25 MV photon beams; a maximum deviation of 1.3% is observed at 18 MV. Calculated products of output ratios and off-axis ratios derived from films are compared with measurements for asymmetric square, rectangular and irregular fields, and agree mostly within 1% for all energies checked; maximum deviations of 1.3 and 1.6% are observed for 6 and 18 MV photon beams.Keywords
This publication has 31 references indexed in Scilit:
- Dosimetric comparison of an integrated multileaf-collimator versus a conventional collimatorPhysics in Medicine & Biology, 1997
- Measurements of basic parameters in wedged high-energy photon beams using a mini-phantomPhysics in Medicine & Biology, 1997
- Dosimetry of high energy photon and electron beams with CEA filmsMedical Physics, 1996
- A dosimetric investigation of scatter conditions for dual asymmetric collimators in open fieldsMedical Physics, 1995
- On the field-size dependence of relative output from a linear acceleratorMedical Physics, 1992
- Clinical dosimetry for implementation of a multileaf collimatorMedical Physics, 1992
- Design Principles and Clinical Possibilities with A New Generation of Radiation Therapy Equipment: A ReviewActa Oncologica, 1987
- Off-center ratios for three-dimensional dose calculationsMedical Physics, 1986
- FILM DOSIMETRY OF HIGH‐ENERGY ELECTRONSAnnals of the New York Academy of Sciences, 1969
- A Note on the Calculation of Dose in X-ray FieldsThe British Journal of Radiology, 1950