Evaluation of Factors Affecting CGMS Calibration
- 1 June 2006
- journal article
- research article
- Published by Mary Ann Liebert Inc in Diabetes Technology & Therapeutics
- Vol. 8 (3) , 318-325
- https://doi.org/10.1089/dia.2006.8.318
Abstract
Background: The optimal number/timing of calibrations entered into the CGMS® (Medtronic MiniMed, Northridge, CA) continuous glucose monitoring system have not been previously described. Methods: Fifty subjects with Type 1 diabetes mellitus (10–18 years old) were hospitalized in a clinical research center for ∼ 24 h on two separate days. CGMS and OneTouch® Ultra® meter (LifeScan, Milpitas, CA) data were obtained. The CGMS was retrospectively recalibrated using the Ultra data varying the number and timing of calibrations. Resulting CGMS values were compared against laboratory reference values. Results: There was a modest improvement in accuracy with increasing number of calibrations. The median relative absolute deviation (RAD) was 14%, 15%, 13%, and 13% when using three, four, five, and seven calibration values, respectively (P < 0.001). Corresponding percentages of CGMS–reference pairs meeting the International Organisation for Standardisation criteria were 66%, 67%, 71%, and 72% (P < 0.001). Nighttime accuracy improved when daytime calibrations (pre-lunch and pre-dinner) were removed leaving only two calibrations at 9 p.m. and 6 a.m. (median difference, −2 vs. −9 mg/dL, P < 0.001; median RAD, 12% vs. 15%, P = 0.001). Accuracy was better on visits where the average absolute rate of glucose change at the times of calibration was lower. On visits with average absolute rates P = 0.05). Conclusions: Although accuracy is slightly improved with more calibrations, the timing of the calibrations appears more important. Modifying the algorithm to put less weight on daytime calibrations for nighttime values and calibrating during times of relative glucose stability may have greater impact on accuracy.Keywords
This publication has 10 references indexed in Scilit:
- Impact of Exercise on Overnight Glycemic Control in Children with Type 1 Diabetes MellitusThe Journal of Pediatrics, 2005
- Accuracy of the Modified Continuous Glucose Monitoring (CGMS®) Sensor in an Outpatient Setting: Results from a Diabetes Research in Children Network (DirecNet) StudyDiabetes Technology & Therapeutics, 2005
- Timing of Changes in Interstitial and Venous Blood Glucose Measured With a Continuous Subcutaneous Glucose SensorDiabetes, 2003
- The Accuracy of the CGMS™in Children with Type 1 Diabetes: Results of the Diabetes Research in Children Network (DirecNet) Accuracy StudyDiabetes Technology & Therapeutics, 2003
- A Continuous Glucose Sensor Based on Wired Enzyme™ Technology - Results from a 3-Day Trial in Patients with Type 1 DiabetesDiabetes Technology & Therapeutics, 2003
- Spurious Reporting of Nocturnal Hypoglycemia by CGMS in Patients With Tightly Controlled Type 1 DiabetesDiabetes Care, 2002
- Do Sensor Glucose Levels Accurately Predict Plasma Glucose Concentrations During Hypoglycemia and Hyperinsulinemia?Diabetes Care, 2002
- Calibration of a subcutaneous amperometric glucose sensor implanted for 7 days in diabetic patientsBiosensors and Bioelectronics, 2001
- Evaluation and comparison of 10 glucose methods and the reference method recommended in the proposed product class standard (1974).Clinical Chemistry, 1977