Five‐Year Performance of the Medtronic 6971 Polyurethane Endocardial Electrode
- 1 November 1986
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 9 (6) , 1173-1180
- https://doi.org/10.1111/j.1540-8159.1986.tb06689.x
Abstract
There is a considerable debate over the long-term performance of polyurethane electrodes. Observation to date has demonstrated surface cracking and some clinical failures in patients. Since 1978 we have implanted about 82 6971 Unipolar Medtronic electrodes in the ventricle. In 33 patients, lead integrity was assessed by chronic thresholds determined by decreasing pulse width and pulse amplitude. Sensing functions were assessed by electrocardiographic rhythm analysis. At a follow-up between 7 and 67 months, chronic thresholds at 2.50 volts were 0.08 +/- 0.04 milliseconds in 26 patients with Medtronic Pulse generators (Models 8423, 5985, and 7000). In two patients with similar units, no loss of capture was seen even at 0.05 milliseconds. Three patients, one with Cordis Unit (233F) showed loss of capture at 0.20 milliseconds at 2.00 MA, the other with a Pacesetter unit (255-6) showed loss of capture at 0.20 milliseconds, at 2.50 volts. One patient with Intermedics unit (283) lost capture at 0.07 milliseconds at 2.70 volts. Insulation breaks seen in two patients were demonstrated by pectoral stimulation and pacemaker oversensing. In addition, 24-hour long-term electrocardiographic monitoring was performed in 22 patients between 23 to 70 months. Appropriate pacemaker function was seen except in one patient who demonstrated oversensing. Interruption in insulation was demonstrated at the ligature site at exploration. Thus, in this series of patients who were paced in the ventricle by the 6971 Medtronic electrode, only two patients have demonstrated insulation failure. The incidence of insulation break in this polyurethane unipolar electrode is uncommon and occurs at further stress points.Keywords
This publication has 16 references indexed in Scilit:
- Bipolar Tined Polyurethane Ventricular Lead: A Four‐Year ExperiencePacing and Clinical Electrophysiology, 1985
- Sixteen Failures in a Single Model of Bipolar Polyurethane‐insulated Ventricular Pacing Lead: A 44‐Month ExperiencePacing and Clinical Electrophysiology, 1984
- Insulation Failure with Bipolar Polyurethane Pacing LeadsPacing and Clinical Electrophysiology, 1984
- Difficulty of extraction of chronically implanted tined ventricular endocardial leadsJournal of the American College of Cardiology, 1984
- To the EditorPacing and Clinical Electrophysiology, 1984
- Self‐Inhibition of an AV Sequential Demand (DVI) Pulse Generator Due to Polyurethane Lead Insulation DisruptionPacing and Clinical Electrophysiology, 1983
- False Inhibition of an Atrial Demand Pacemaker Caused by an Insulation Defect in a Polyurethane LeadPacing and Clinical Electrophysiology, 1983
- Polyurethane LeadsPacing and Clinical Electrophysiology, 1983
- The Polyurethane ControversyPacing and Clinical Electrophysiology, 1983