Ultrafast Differential Ion Mobility Spectrometry at Extreme Electric Fields Coupled to Mass Spectrometry
- 26 August 2009
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 81 (19) , 8048-8053
- https://doi.org/10.1021/ac901479e
Abstract
Microchip-based field asymmetric waveform ion mobility spectrometry (FAIMS) analyzers featuring a grid of 35 μm-wide channels have allowed electric field intensity (E) over 60 kV/cm, or about twice that in previous devices with >0.5 mm gaps. Since the separation speed scales as E4 to E6, these chips filter ions in just ∼20 μs (or ∼100−10 000 times faster than “macroscopic” designs), although with reduced resolution. Here we report integration of these chips into electrospray ionization (ESI) mass spectrometry, with ESI coupled to FAIMS via a curtain plate/orifice interface with edgewise ion injection into the gap. Adjusting gas flows in the system permits control of ion residence time in FAIMS, which affects resolving power independently of ion desolvation after the ESI source. The results agree with a priori simulations and scaling rules. Applications illustrated include analyses of amino acids and peptides. Because of limited resolving power, the present FAIMS units are more suitable for distinguishing compound classes than individual species. In particular, peptides separate from many other classes, including PEGs that are commonly encountered in proteomic analyses. In practical analyses with realistic time constraints, the effective separation power of present FAIMS may approach that of “macroscopic” systems.This publication has 43 references indexed in Scilit:
- Improving FAIMS sensitivity using a planar geometry with slit interfacesJournal of the American Society for Mass Spectrometry, 2009
- Ultrafast Differential Ion Mobility Spectrometry at Extreme Electric Fields in Multichannel MicrochipsAnalytical Chemistry, 2009
- IMS/MS: its time has comeAnalytical Chemistry, 2008
- Control of Ion Distortion in Field Asymmetric Waveform Ion Mobility Spectrometry via Variation of Dispersion Field and Gas TemperatureAnalytical Chemistry, 2008
- Assessing the Dynamic Range and Peak Capacity of Nanoflow LC−FAIMS−MS on an Ion Trap Mass Spectrometer for ProteomicsAnalytical Chemistry, 2008
- Dynamically Multiplexed Ion Mobility Time-of-Flight Mass SpectrometryAnalytical Chemistry, 2008
- Simultaneous fragmentation of multiple ions using IMS drift time dependent collision energiesJournal of the American Society for Mass Spectrometry, 2008
- Ion mobility spectrometry—mass spectrometry performance using electrodynamic ion funnels and elevated drift gas pressuresPublished by American Chemical Society (ACS) ,2007
- High-Resolution Field Asymmetric Waveform Ion Mobility Spectrometry Using New Planar Geometry AnalyzersAnalytical Chemistry, 2006
- Characterizing the Structures and Folding of Free Proteins Using 2-D Gas-Phase Separations: Observation of Multiple Unfolded ConformersAnalytical Chemistry, 2006