Method for the calibration of atomic force microscope cantilevers
- 1 July 1995
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 66 (7) , 3789-3798
- https://doi.org/10.1063/1.1145439
Abstract
The determination of the spring constants of atomic force microscope (AFM) cantilevers is of fundamental importance to users of the AFM. In this paper, a fast and nondestructive method for the evaluation of the spring constant which relies solely on the determination of the unloaded resonant frequency of the cantilever, a knowledge of its density or mass, and its dimensions is proposed. This is in contrast to the method of Cleveland et al. [Rev. Sci. Instrum. 64, 403 (1993)], which requires the attachment of masses to the cantilever in the determination of the spring constant. A number of factors which can influence the resonant frequency are examined, in particular (i) gold coating, which can result in a dramatic variation in the resonant frequency, for which a theoretical account is presented and (ii) air damping which, it is found, leads to a shift of ∼4% in the resonant frequency down on its value in a vacuum. Furthermore, the point of load on the cantilever is found to be extremely important, since a small variation in the load point can lead to a dramatic variation in the spring constant. Theoretical results that account for this variation, which, it is believed will be of great practical value to the users of the AFM, are given.Keywords
This publication has 24 references indexed in Scilit:
- Resonance response of scanning force microscopy cantileversReview of Scientific Instruments, 1994
- Lateral, normal, and longitudinal spring constants of atomic force microscopy cantileversReview of Scientific Instruments, 1994
- Experimental Determination of Spring Constants in Atomic Force MicroscopyLangmuir, 1994
- Theoretical analysis of the static deflection of plates for atomic force microscope applicationsJournal of Applied Physics, 1993
- Calibration of atomic-force microscope tipsReview of Scientific Instruments, 1993
- Direct measurement of interaction forces between colloidal particles using the scanning force microscopeLangmuir, 1993
- A nondestructive method for determining the spring constant of cantilevers for scanning force microscopyReview of Scientific Instruments, 1993
- Atomic force microscopyProgress in Surface Science, 1992
- Microfabrication of cantilever styli for the atomic force microscopeJournal of Vacuum Science & Technology A, 1990
- Young’s modulus measurements of thin films using micromechanicsJournal of Applied Physics, 1979