A hip joint simulator study using simplified loading and motion cycles generating physiological wear paths and rates
- 1 June 1999
- journal article
- research article
- Published by SAGE Publications in Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
- Vol. 213 (6) , 455-467
- https://doi.org/10.1243/0954411991535068
Abstract
In some designs of hip joint simulator the cost of building a highly complex machine has been offset with the requirement for a large number of test stations. The application of the wear results generated by these machines depends on their ability to reproduce physiological wear rates and processes. In this study a hip joint simulator has been shown to reproduce physiological wear using only one load vector and two degrees of motion with simplified input cycles. The actual path of points on the femoral head relative to the acetabular cup were calculated and compared for physiological and simplified input cycles. The in vitro wear rates were found to be highly dependent on the shape of these paths and similarities could be drawn between the shape of the physiological paths and the simplified elliptical paths.Keywords
This publication has 30 references indexed in Scilit:
- Wear and Morphology of Ultra-High Molecular Weight Polyethylene Wear Particles from Total Hip ReplacementsProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1996
- The Importance of Multidirectional Motion on the Wear of PolyethyleneProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1996
- Mechanistic and Morphological Origins of Ultra-High Molecular Weight Polyethylene Wear Debris in Total Joint Replacement ProsthesesProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1996
- Ultra-High Molecular Weight Polyethylene Wear Debris Generated in Vivo and in Laboratory Tests; the Influence of Counterface RoughnessProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1996
- Comparison of the size and morphology of UHMWPE wear debris produced by a hip joint simulator under serum and water lubricated conditionsBiomaterials, 1996
- Tissue ingrowth and differentiation in the bone-harvest chamber in the presence of cobalt-chromium-alloy and high-density-polyethylene particles.Journal of Bone and Joint Surgery, 1995
- The effect of transfer film and surface roughness on the wear of lubricated ultra-high molecular weight polyethyleneClinical Materials, 1993
- Wear studies on prosthetic materials using the pin-on-disc machineBiomaterials, 1982
- Wear of Artificial Joint Materials IVEngineering in Medicine, 1981
- Reactions of the articular capsule to wear products of artificial joint prosthesesJournal of Biomedical Materials Research, 1977