Load transfer through a hydroxyapatite‐coated canine hip implant
- 1 December 1994
- journal article
- research article
- Published by Wiley in Journal of Applied Biomaterials
- Vol. 5 (4) , 293-306
- https://doi.org/10.1002/jab.770050404
Abstract
Strain transfer near hydroxyapatite (HA) coated canine hip implants was examined using simulated anatomical loading based on in vivo strain measurements. Strain changes near implants relative to intact control values were in excess of 100% for transverse and principal strains for zero time period (immediate postimplant) specimens. They were generally smaller (100% or less) for axial, transverse, and principal minimums in the same locations for 4 months postimplantation specimens.Cortical bone loss occurred in all implanted femora. The most extensive loss, up to 47%, occurred adjacent to the proximal section of the implant. Extensive trabecular bone formation, over 300% in some regions of each femur, was noted in all implanted femora.Backscattered electron imaging along the HA‐coated sections of the implants showed extensive bone bonded to the coating. Normal light and UV light micrographs showed direct bone apposition to the implant surfaces and extensive bone formation in all test animals. Microscopy revealed no evidence of any soft tissue layer between the implant and bone. Bone was typically found in direct contact with the implant surface.Histomorphometry indicated that bone formation rates in the implanted femora were elevated, up to 850%, relative to controls. Fewer formation sites were noted on the posterior and lateral (in two cases zero sites). Mineral apposition rates (MAR) from two of the dogs were slightly elevated (from 110–113%) in the implanted femora relative to controls and depressed (to about 83% of controls) in a third. © 1994 John Wiley & Sons, Inc.Keywords
This publication has 26 references indexed in Scilit:
- Bone Remodeling and in Vivo Strain Analysis of Intact and Implanted Greyhound Proximal FemoraJournal of Investigative Surgery, 1994
- In Vivo Strain Analysis of the Greyhound Femoral DiaphysisJournal of Investigative Surgery, 1992
- Alterations in femoral and acetabular bone strains immediately following cementless total hip arthroplasty: An in vitro canine studyJournal of Orthopaedic Research, 1991
- Retrieval Analysis of a Hydroxyapatite-Coated Hip ProsthesisPublished by Wolters Kluwer Health ,1991
- Hydroxyapatite-Coated Total Hip Prostheses Two-Year Clinical and Roentgenographic Results of 100 CasesClinical Orthopaedics and Related Research, 1990
- A Long-Term in Vivo Bone Strain Measurement DeviceJournal of Investigative Surgery, 1989
- A Canine Composite Femoral StemPublished by Wolters Kluwer Health ,1988
- Die biologische Leistung der Hydroxylapatitkeramik-Beschichtung auf dem Femurschaft einer Titanendoprothese — erste histologische Auswertung eines Humanexplantats - The Biological Behavior of the Hydroxyapatite Ceramic Coating on a Titanium Stern of a Hip Prosthesis - the First Histological Evaluating of Human Autopsy MaterialBiomedizinische Technik/Biomedical Engineering, 1987
- A Polymethyl Methacrylate Method for Large Specimens of Mineralized Bone with ImplantsStain Technology, 1987
- The effect of proximally and fully porous‐coated canine hip stem design on bone modelingJournal of Orthopaedic Research, 1987