Measurement of Cardiac Output in Man by Injection of Dye at a Constant Rate Into the Right Ventricle or Pulmonary Artery
- 1 May 1955
- journal article
- research article
- Published by American Physiological Society in Journal of Applied Physiology
- Vol. 7 (6) , 629-638
- https://doi.org/10.1152/jappl.1955.7.6.629
Abstract
A method is described for the determination of the cardiac output in man by the injection of Evans blue dye (T-1824) at a constant rate into the right ventricle or pulmonary artery with continuous recording of the dye-dilution pattern by a cuvette oximeter connected to the radial artery. The method is simple to carry out and results can be quickly calculated. Evidence is presented which indicates that significant recirculation of dye does not occur prior to the establishment of an equilibrium or near-equilibrium concentration of dye in the radial-artery blood following such injections. Determinations of cardiac output by this constant-rate injection technique, by the Fick method and by the Hamilton method following the sudden single injection of dye were made in 22 studies on 17 subjects. Cardiac output values by the Fick method ranged from 1.7 to 7.7 l/minute/m2. There were no systematic differences between the 3 methods. Standard deviation of differences between the paired Fick and constant-rate dye_injection values was 10%, between sudden single and constant-rate dye-injection values was 16% and between Fick and sudden single dye-injection values was 16%. The average time intervals between these paired determinations were 5, 9 and 6 minutes, respectively. It is concluded that the constant-rate-injection version of the indicator-dilution technique is a reliable method for the measurement of cardiac output in man when the pulmonary artery or right ventricle is used as the injection site.Keywords
This publication has 10 references indexed in Scilit:
- Measurement of Left Ventricular OutputJournal of Applied Physiology, 1954
- Simultaneous Measurement of the Appearance and Disappearance of T-1824 (Evans Blue) in Blood and Tissue After Intravenous Injection in ManJournal of Applied Physiology, 1954
- Limitations of the Continuous Infusion Method for Measuring Cardiac Output by Dye DilutionAmerican Journal of Physiology-Legacy Content, 1953
- A METHOD FOR THE CONTINUOUS RECORDING OF EVANS BLUE DYE CURVES IN ARTERIAL BLOOD, AND ITS APPLICATION TO THE DIAGNOSIS OF CARDIOVASCULAR ABNORMALITIES1951
- COMPARISON OF THE CONSTANT AND INSTANTANEOUS INJECTION TECHNIQUES FOR DETERMINING CARDIAC OUTPUTAmerican Journal of Physiology-Legacy Content, 1949
- Comparison of the Fick and Hamilton Methods for the Determination of Cardiac Output in Man121Scandinavian Journal of Clinical and Laboratory Investigation, 1949
- COMPARISON OF THE FICK AND DYE INJECTION METHODS OF MEASURING THE CARDIAC OUTPUT IN MANAmerican Journal of Physiology-Legacy Content, 1948
- COMPARISON OF THE TIME CONCENTRATION CURVES IN ARTERIAL BLOOD OF DIFFUSIBLE AND NON-DIFFUSIBLE SUBSTANCES WHEN INJECTED AT A CONSTANT RATE AND WHEN INJECTED INSTANTANEOUSLYAmerican Journal of Physiology-Legacy Content, 1947
- THE EFFECT OF POSITIVE AND NEGATIVE INTRATHORACIC PRESSURE ON CARDIAC OUTPUT AND VENOUS PRESSURE IN THE DOGAmerican Journal of Physiology-Legacy Content, 1944
- CARDIAC OUTPUT AND TOTAL PERIPHERAL RESISTANCE MEASUREMENTS IN EXPERIMENTAL DOGSAmerican Journal of Physiology-Legacy Content, 1944