Uncertainty analysis in polymer gel dosimetry
- 20 October 1999
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 44 (11) , N243-N246
- https://doi.org/10.1088/0031-9155/44/11/402
Abstract
Verification of advanced radiotherapy treatment modalities requires measurement of three-dimensional absorbed dose distributions with high spatial resolution and precision. Polymer gel dosimeters combined with magnetic resonance imaging may be able to fulfil this requirement. However, verification requires that the uncertainty in the dosimeter is well known. One method of estimating the overall uncertainty in polymer gel dosimeters involves the propagation of the uncertainty in the R2 (nuclear magnetic resonance relaxation rate) map and the uncertainties in the calibration data. This work shows that using this method with current data suggests that the lowest uncertainty currently obtainable is about 3% at 8 Gy and 7% at 2 Gy. Furthermore, the most significant reductions in overall uncertainty will be achieved by reducing the noise in the R2 map.Keywords
This publication has 3 references indexed in Scilit:
- Improving calibration accuracy in gel dosimetryPhysics in Medicine & Biology, 1998
- An investigation into the dosimetry of a nine-field tomotherapy irradiation using BANG-gel dosimetryPhysics in Medicine & Biology, 1998
- Experimental procedure for the manufacture and calibration of polyacrylamide gel (PAG) for magnetic resonance imaging (MRI) radiation dosimetryPhysics in Medicine & Biology, 1998