Repeatability of Layered Corneal Pachymetry with the Artemis Very High-Frequency Digital Ultrasound Arc-Scanner
- 1 September 2010
- journal article
- Published by SLACK, Inc. in Journal of Refractive Surgery
- Vol. 26 (9) , 646-659
- https://doi.org/10.3928/1081597x-20091105-01
Abstract
Purpose: To assess the three-dimensional repeatability of thickness measurements for epithelium, stroma, cornea, flap, and residual stromal bed using the Artemis very high-frequency (VHF) digital ultrasound arc-scanner (ArcScan Inc). Methods: Five consecutive measurements were obtained for 10 eyes of 10 patients 1 year after LASIK using the Artemis VHF digital ultrasound arc-scanner across the central 10-mm diameter of the cornea. Repeatability analysis was performed for thickness measurements for each corneal layer—epithelium, stroma, cornea, flap, and residual stromal bed. The standard deviation of repeated measurements (point-repeatability) was calculated for each measurement location in 0.1-mm steps for the 10×10-mm matrix. The pooled standard deviation of the point-repeatability for each measurement location within the central 1-, 2-, and 3-mm radius was calculated (region-repeatability). The corneal thickness of the baseline scan set was compared to that of subsequent scan sets within the same session and plotted over time to assess any possible hydration effects of the immersion technique. Results: The repeatability at the corneal vertex was 0.58 μ m for epithelium, 1.78 μ m for stroma, 1.68 μ m for cornea, 1.68 μ m for flap, and 2.27 μ m for residual stromal bed. The region-repeatability within the central 1-mm radius was 1.01 μ m for epithelium, 3.44 μ m for stroma, 3.35 μ m for cornea, 2.81 μ m for flap, and 3.97 μ m for residual stromal bed. The mean difference in corneal thickness from the baseline value was within 1.25 μ m for each of the subsequent four scan sets over a 5-minute immersion period. Conclusions: Layered pachymetry of the epithelium, stroma, cornea, flap, and residual stromal bed showed high repeatability with the Artemis VHF digital ultrasound arc-scanner. The high repeatability validates the use of the Artemis for in vivo layered pachymetry.Keywords
This publication has 87 references indexed in Scilit:
- Epithelial Thickness After Hyperopic LASIK: Three-Dimensional Display with Artemis Very High-Frequency Digital UltrasoundJournal of Refractive Surgery, 2010
- Epithelial, Stromal, and Total Corneal Thickness in Keratoconus: Three-dimensional Display With Artemis Very-high Frequency Digital UltrasoundJournal of Refractive Surgery, 2010
- Stromal Thickness in the Normal Cornea: Three-dimensional Display With Artemis Very High-Frequency Digital UltrasoundJournal of Refractive Surgery, 2009
- Epithelial, Stromal, and Corneal Pachymetry Changes during OrthokeratologyOptometry and Vision Science, 2009
- Effect of Corneal Hydration on Ultrasound Velocity and BackscatterUltrasound in Medicine & Biology, 2009
- Central corneal thickness measured by the Orbscan II system, contact ultrasound pachymetry, and the Artemis 2 systemJournal of Cataract & Refractive Surgery, 2008
- Accuracy, repeatability, and reproducibility of Artemis very high-frequency digital ultrasound arc-scan lateral dimension measurementsJournal of Cataract & Refractive Surgery, 2006
- The Repeatability of Corneal and Corneal Epithelial Thickness Measurements Using Optical Coherence TomographyOptometry and Vision Science, 2006
- Visualizing VHF ultrasound of the human corneaIEEE Computer Graphics and Applications, 1999
- High-Frequency Ultrasound Corneal Pachymetry in the Assessment of Corneal Scars for Therapeutic PlanningEye & Contact Lens: Science & Clinical Practice, 1994