Prospective Randomized Comparison of 8‐mm Gold‐Tip, Externally Irrigated‐Tip and 8‐mm Platinum‐Iridium Tip Catheters for Cavotricuspid Isthmus Ablation
- 30 May 2007
- journal article
- research article
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 18 (7) , 709-713
- https://doi.org/10.1111/j.1540-8167.2007.00861.x
Abstract
Radiofrequency (RF) ablation of the cavotricuspid isthmus (CTI) can be performed using different types of ablation catheter. Gold tip electrodes have the theoretical advantage of creating bigger lesions than standard platinum-iridium electrode. This prospective, randomized study compares the clinical efficacy of 8-mm gold tip catheter, externally irrigated and 8-mm platinum-iridium tip (Pt tip) catheters. Sixty consecutive patients (51 men, 60 +/- 10 years) undergoing de novo CTI ablation for documented typical atrial flutter were randomized to one of the following ablation catheters: 8-mm gold tip catheter, an externally irrigated-tip (Irr. tip) catheter, or an 8-mm Pt tip catheter. The procedural endpoint was achievement of bidirectional isthmus conduction block with < or = 20 minutes of RF energy application. The latter was achieved equally with the 3 catheters (95% for gold tip, 100% for irrigated tip, 95% for Pt tip) and the durations of RF (10 +/- 6, 10 +/- 4, 13 +/- 8 minutes), fluoroscopy (12 +/- 6, 12 +/- 7, 15 +/- 12 minutes) and the procedure (34 +/- 23, 38 +/- 24, 40 +/- 30 minutes) were similar in all groups. The maximal targeted power could not be reached in at least one location in 40% of patients with gold tip and in 35% of patients with Pt tip catheters whereas it was always achieved with an Irr. tip catheter (P = 0.003, P = 0.008). The reduction in impedance during RF delivery was greater with Irr. tip (11 +/- 7 ohms) than with gold (7 +/- 4 ohms, P = 0.02) or Pt tip (5 +/- 3 ohms, P = 0.001) catheters. This study demonstrates equivalent efficacies of gold, platinum-iridium and externally Irr. tip catheters for successful de novo ablation of the CTI.Keywords
This publication has 29 references indexed in Scilit:
- Cavotricuspid isthmus angiography predicts atrial flutter ablation efficacy in 281 patients randomized between 8 mm- and externally irrigated-tip catheterEuropean Heart Journal, 2006
- LETTERS TO THE EDITOR: 3Journal of Cardiovascular Electrophysiology, 2005
- Comparison of the Efficacy of Cooled-Tip and 8-mm-Tip Catheters for Radiofrequency Catheter Ablation of the Cavotricuspid Isthmus: A Meta-AnalysisPacing and Clinical Electrophysiology, 2005
- Gold‐Tip Electrodes—A New “Deep Lesion” Technology for Catheter Ablation? In Vitro Comparison of a Gold Alloy Versus Platinum–Iridium Tip Electrode Ablation CatheterJournal of Cardiovascular Electrophysiology, 2005
- Pattern of Isthmus Conduction Recovery Using Open Cooled and Solid Large‐Tip Catheters for Radiofrequency Ablation of Typical Atrial FlutterJournal of Cardiovascular Electrophysiology, 2004
- Effect of Isthmus Anatomy and Ablation Catheter on Radiofrequency Catheter Ablation of the Cavotricuspid IsthmusCirculation, 2004
- Prospective randomised comparison of irrigated-tip and large-tip catheter ablation of cavotricuspid isthmus-dependent atrial flutter*1Published by Oxford University Press (OUP) ,2004
- Why a Large Tip Electrode Makes a Deeper Radiofrequency Lesion:Journal of Cardiovascular Electrophysiology, 1998
- Comparison of Gold Versus Platinum Electrodes on Myocardial Lesion Size Using Radiofrequency EnergyPacing and Clinical Electrophysiology, 1996